User manual
PONSEL
Référence : NOTICE_ACTEON_2050_v001
V : 002
25/67
.
2) The second step consists in entering the sludge sample dry weight obtained after a variable delay in order to
complete the SS sensor calibration, which means calculating and recording the new slope.
This step is not affected by the state of the optical sensor, which can be disconnected from the device, connected
and in the air or preferably connected and already in the tank. In the latter case, at the end of this step and
immediately following calibration confirmation and automatic return to MEASURE mode, the g/l value
displayed will be that of the SS content of the sludge measured at the sensor's location in the tank.
CALIBRATION
SENSORS CALIBRATION
SS SENSOR
Gain adjust
Immediate calibration
Content of SS after analysis
Theoretical calibration
TEMPERATURE SENSOR
Gain adjust
Complete calibration
Theoretical calibration
ENTER: Validate the choice
Use the
and
keys to select
the type of calibration and the
probe to be calibrated.
Select the
Content of SS after analysis
menu and press
ENTER
.
Comment: The measurement window is displayed again once the calibration has
been correctly carried out
The displayed MEASURE g/l value remains unchanged. The previous slope was not
overwritten and remains displayed (until end of 2nd step), and the SS-meter will give
another useable measurement: this is still attributed to the previous calibration slope
coefficient and not the current one.
20.00
°C
-24 hrs
2.31
MEASURE
g/l
INFO
CONF
HIST
CAL
LOG
MEAS
Mode1 R1
R2
ACTEON 2050-20 Turbidimeter
g/l
1
3
DRY WEIGHT
!
Indicates the device awaits dry
weight input following sludge
sample analysis