User manual
PONSEL
Référence : NOTICE_ACTEON_2050_v001
V : 002
51/67
9.6 Adjusting relay outputs
The 2 relays can be configured in 3 different modes:
- Mode 1 is used to configure relay R1 on an SS measurement value and relay R2 on a temperature measurement
value.
- Mode 2 is used to configure relays R1 and R2 on two SS measurement values.
- Mode 3 is used to configure:
- relay R1 on two SS measurement thresholds in regulation mode (forced start-up and shutdown delays
can be configured).
- relay R2 on a temperature measurement threshold in alarm mode.
The 2 relays can be configured with the following attributes:
-
Alarm threshold: Threshold trigger value
-
Hysteresis: Hysteresis value for relay switching (prevents relay hunting)
-
Triggering direction:
Above: Means the relay contact is closed above the alarm threshold.
Below: Means the relay contact is open below the alarm threshold.
-
Forced start-up and shutdown delay (only in mode 3)
alarme
Use the
and
keys to select
the configuration required.
Select the
Relay adjustment
menu and press ENTER.
Use the
and
keys to select the
configuration required.
CONFIG
Configuration of the relays
mode:
- MODE 1: Alarms adjustment
1 threshold on SS
1 threshold on the temperature
- MODE 2: Regulation 2 relays
2 thresholds on SS
- MODE 3: Regulation 1 relay
2 thresholds on SS
1 threshold on the temperature
ESC: Escape menu
Enter: Validate the choice
CONFIG
WHAT DO YOU WANT TO DO?
Averaging sensor response
Sludge slope coefficient reminder
Sensor offset adjustment
Graphic screen
4-20mA outputs adjustment
Relay adjustment
Measure units
Language
Factory default value reminder
ESC: Escape menu