6.1.4 Returning to SS measurement theoretical calibration:
Theoretical calibration is carried out using SS sensor theoretical coefficients.
20.00
°C
-24 hrs
2.31
MEASURE
g/l
INFO
CONF
HIST
CAL
LOG
MEAS
Mode1 R1
R2
ACTEON 2050-20 Turbidimeter
g/l
1
3
Use the
and
keys to select
the type of calibration and the
probe to be calibrated.
Select the
Theoretical calibration
menu and
press ENTER.
THEOR CAL
!
Indicates that the device has
been calibrated with the
theoretical coefficients.
Use the
and
keys to select the
procedure confirmation.
CALIBRATION
THEORETICAL CALIBRATION
Esc: Cancel the procedure
Enter: Validate the choice
ATTENTION, you will erase your
calibration coefficients to return to
the theoretical coefficients!
Yes
No
Select
Yes
and press ENTER to return to the
theoretical coefficients.
ENTER : Validate the choice
CALIBRATION
SENSORS CALIBRATION
SS SENSOR
Gain adjust
Immediate calibration
Differed calibration
Theoretical calibration
TEMPERATURE SENSOR
Gain adjust
Complete calibration
Theoretical calibration
ENTER: Validate the choice