5-2
5 Troubleshooting
5.2
Problems with the Interferometer
No light is
emitted
If no light is being emitted from the interferometer, please check if the supply
cable (Supply) between the controller and the interferometer has been
installed and secured correctly.
Problems with
the piezo
Set the measurement distance (approx. 40 mm), by slowly changing the
distance between the interferometer and the object until the interferences are
visible in the software. If you now move the piezo position in the software and
the interference pattern is not changing, please check the following:
1. Has the piezo cable (Piezo) between the controller and the interferometer
been installed and secured correctly with the screws on the plug?
2. Has the piezo control cable (Piezo Control) between the PC and the
controller been installed and secured correctly?