3-10
3 First Steps
12
COM 1 serial interface (9-pin Sub-D plug)
This connection is not used.
13
Monitor connection (DisplayPort jack)
This connection is not used.
14 KEYBOARD
connection (6-pin circular jack)
15 MOUSE
connection (6-pin circular jack)
This connection is not used.
16
Mains switch
17
Mains connection (socket for standard power cord)
3.3
Assembly
CAUTION!
Risk of injury caused by heavy system components !
There is the risk of injury to hands and feet when carrying and lifting heavy
system components.
» Only lift system components in pairs or using a suitable lifting device.
» Wear safety shoes.
» Wear protective gloves.
ATTENTION!
Damage of the device caused by insufficient opportunity for air
circulation!
Heat is created inside the device during operation. If the heat can not exhaust
from the housing it leads to heat accumulation. This will overheat the
electronic parts inside the device and may result in a defect.
» Always keep free all air inlets and outlets of the device.
» Keep at least 50 mm distance to the wall if required when you set up the
device!
» If the devices are in a system cabinet: Remove all covers of the system
cabinet before switching on the devices.
» If you notice a cooling fan failure, switch off the device immediately.
ATTENTION!
Misalignment of the optics by shock!
The optics of the sensor head are highly sensitive. The optics can get
misaligned by shocks or vibrations.
» Protect the unpacked device from being shaken.
INFORMATION
Before starting to assemble the instrument, make sure that all system
components are in perfect condition! Check the devices for external damages
such as scratches or the like. In the case of a wrong delivery, damage or
missing parts, please inform your local Polytec representative immediately,
stating the serial number of the devices.