3-11
3 First Steps
INFORMATION
Keep the assembly instructions for all components in a safe place.
3.3.1
Assembling or Exchanging the Desiccant Cartridge
ATTENTION!
Damage caused by moisture!
If you do not exchange the desiccant cartridge, moisture can condense on the
optical components in the interferometer and can distort the measurement
result. In an extreme case, the instrument could be damaged !
» Exchange the desiccant cartridge when it is saturated
INFORMATION
Keep the see-through sleeve of the desiccant cartridge you have installed to
replace the desiccant cartridge subsequently.
INFORMATION
If you pack the instrument for long transport where it may be subjected to
greatly fluctuating climatic conditions (e.g. by air freight), please remove the
desiccant cartridge which has been installed and simply enclose the
desiccant cartridge in its original packaging with the instrument. Make sure
that the opening on the back of the interferometer is sealed with protective
film.
The moisture indicator on the desiccant cartridge shows you when it needs to
be changed. The moisture indicator changes color once the desiccant
cartridge is saturated and can not absorb any more moisture. Because of this,
regularly monitor the condition of the moisture indicator and exchange the
desiccant cartridge if necessary.
To assemble or resp. exchange the desiccant cartridge, proceed as follows:
1. If none desiccant cartridge is assembled yet: At the back of the
interferometer, pull the protective foil from the bracket of the desiccant
cartridge.
2. If a desiccant cartridge is assembled already: At the back of the
interferometer, use the back of the transparent packing of a desiccant
cartridge to turn the installed desiccant cartridge counterclockwise from
the bracket.
3. Use a transparent packing of a desiccant cartridge to twist another
desiccant cartridge counterclockwise out of its transparent packing.
4. Using the transparent packing, screw the (new) desiccant cartridge
clockwise into the holder on the back of the interferometer (refer to
5. Monitor the moisture indicator on the desiccant cartridge.