A-2
A Optional Accessories
A.1.2
Functional Test
For an initial functional test with the geometry scan unit, proceed as follows:
1. Carry out a functional test with the system as described in
and in the software manual.
2. Make a distance measurement (
Geometry Point
) as described in your
software manual.
3. Check the determined stand-off distance (distance between scanning
head and scan point), e.g. with a measuring tape.
If the functional test has been successful you can now make measurements
with the system and the geometry scan unit as described in
4.3 and your software manual.
If your system does not perform as described above, read through the
information on fault diagnosis provided in
6 and, if necessary,
contact your local Polytec representative.
A.1.3
Technical Specifications
Laser
< 1.5 m
Cover the retro-reflective film in any
case. With optically poorly reflective
surfaces treat the object with white
paper and carry out the distance
measurement.
≥
1.5 m
Treat the object with retro-reflective film
and carry out the distance measurement
on the retro-reflective film.
Object under investigation
Distance to the
object
Precautions
Laser type:
Laser diode
Wavelength:
650 … 690 nm
Laser class:
2
Laser power:
< 1 mW
Accuracy of the distance
measurement:
typ. ± 2.5 mm, depends on the signal level