3-3
3 First Steps
Connecting
cables
As a general rule, the instrument may not be switched on until all connecting
cables have been connected up. Make sure that all connections are
connected properly and firmly.
Protect the connecting cables from mechanical damage and high
temperatures. The bending radius of the main cable (Umbilical) may not be
less than 120 mm. We recommend a minimal bending radius of 150 mm, in
order to keep the strain of the main cable as low as possible. The bending
radius of the other cables may not be less than 50 mm.
Mains
connection
The mains voltage inputs of the instruments are designed as wide range
inputs and can be connected to all mains voltages with nominal values in the
range from 100 V to 240 V.
Warming-up
The helium-neon laser will take a while to reach the optimal operating
temperature after it has been switched on. The instrument does not reach its
optimal metrological properties until a warm-up period of approx. 30 minutes.
For measurement tasks with high angular stability requirements we
recommend a warm-up period of 120 minutes. This applies especially when
you are working with imported 3D geometries.
Cleaning
Clean the housing surfaces with mild detergent or disinfectant solutions. Do
not use organic solvents !
Switch off the instrument before cleaning it.
Cleaning
scanning head
Only use bellows to remove dirt such as dust on the scanner mirror.
Switch off the instrument before cleaning it.
Installation of
other
components
Hardware or software components which do not belong to the system can
damage the system or adversely affect the way the Polytec software
functions. Using them will result in the loss of warranty. If you want to install
such components, contact Polytec.
Opening the
instruments
Tampering with the instruments in any way is not necessary when using the
equipment as intended and will invalidate the warranty. Exchanging or
retrospectively installing subassemblies may only be carried out by
authorized service personnel of Polytec.