7-12
7 Technical Specifications
7.5.2
Optics
Figure 7.2:
Depth of focus subject to the stand-off distance
Focal length (nominal):
70 mm
Minimum stand-off distance:
125 mm
Stand-off distance
1
1
Measured from the front of the scanning head
Spot diameter
(1 / e
2
, typ.)
Depth of focus
(typ.)
2
2
For 3 dB signal reduction
Aperture diameter
3
3
At the front of the scanning head
[mm]
[µm]
[mm]
[mm]
141
25
0.48
5.8
345
43
1.4
8.4
549
60
2.8
9.4
753
78
4.6
10
1 000
98
7.5
10.4
With each additional meter stand-off distance 85µm are added to the spot diameter.