8
Table 3.1 continued
3.21 BDAB supply voltage
3.6 (minus 0.6; +0.7) V
3.22 BDAB corresponds to IEC 61000-6-2:2005, IEC 61000-6-3:2006, EN 61326-1:2005
electromagnetic compatibility standards:
- is resistant to industrial frequency radiofrequency magnetic fields of 400 A/m voltage,
performance criterion А;
- is resistant to radiofrequency magnetic fields, test level 4 (30 V/m) within the frequency range
80 – 1000 MHz and 800 – 960 MHz, and 1.4 – 2.5 GHz (when there is electromagnetic emission of
digital radiophones), performance criterion А;
- is resistant to electrostatic discharges, test level 3 (air discharge of 8 kV voltage, contact discharge
of 6 kV voltage), performance criterion В;
- is resistant to is resistant to electromagnetic interference caused by radio frequency
electromagnetic fields, test level 3, performance criterion А;
- corresponds to EN 55022:2010 (B class) requirements by the level of emitted radio-noise
3.23 BDAB withstands falling onto a concrete floor (height 0.7m)
3.24 Maximum weight
0.45 kg
3.25 Overall dimensions
71х45х130 mm
3.26 Reliability figures:
- mean life before failure, no less than
20 000 hours;
- mean service life, no less than
10 years;
- mean time to recovery, no more than
60 min
For additional information on BDAB contact the manufacturer or visit www.polimaster.com