Table 3.1 continued 

3.11 BDAB enables the user to set and constantly monitor the threshold level of 

- and 

- radiation 

fluxes  through  the  whole  measurement  range  and  record  the  set  threshold  level  of 

-  and 

radiation density values in the instrument’s non-volatile memory 

3.12  Spectroscopic  Radiation  Monitor  РM1403  /  Display  and  Control  Unit  PM1403  produces 
discontinuous sound and light alarms and the analogue scale is being filled if set threshold of 

- and 

- radiation flux density is reached/exceeded in the measurement mode of  

- and 

- radiation flux 

density  

3.13 Limits of permissible additional relative error 
of  DER  measurement  of 

-  and 

-  radiation  must 

be no more than: 

- when ambient air changes from normal to  -20°С 
and from normal to + 50 °С 
- at ambient air relative humidity 95 % at 35 °С 
- when supply voltage changes from nominal to 
extreme values during measurement of 

- and 

radiation flux density  
-  under  exposure  of  magnetic  field  strength 
400 A/m  

under 

exposure 

of 

radiofrequency 

electromagnetic fields 

 
 
 
 

 10 %; 

 10 %; 

 
 

 5 % limits of permissible main error; 

 

10 %; 

 

 10 % 

3.14 Limits of permissible additional relative error of 

- radiation flux density measurement don’t 

exceed 30 % under exposure of external gamma radiation field of 200 μR/h exposure rate and don’t 
exceed 15 % under exposure of external 

-radiation 

90

Sr+

90

Y source producing external radiation no 

less than 3.0

10

3

 s

-1

, placed at the distance (50 

 3) mm from the sensitive surface of BDAB; 

3.15 Limits of permissible additional relative error of measurement of 

-particles don’t exceed 15 % 

under exposure of external gamma- radiation field up to 0.5 mR/h exposure dose rate; 
3.16 Unstable BDAB readings when measuring  

and 

-radiation flux density during continuous 

operation (8 hours), no more than 

5 % 

3.17 Operation conditions: 
- ambient air temperature range from -20 to +50 °С; 
- relative humidity of ambient air up to 95 % at 35 

С; 

- atmospheric pressure from 84 to 106.7 kPa 
3.18 BDAB is resistant to: 

-  sinusoidal  vibration  within5-500  Hz  frequency  range  and  shifting  amplitude  for  frequencies 

lower than transition frequency 0.075 mm; 
- blows with 100 m/s

acceleration, shock pulse duration 2-50 ms, blows repetition rate 60 -180 

blows per minute. 
3.19 Protection level of the BDAB case is IP64 
3.20 Information exchange with Spectroscopic 
Radiation Monitor РM1403 / Display and Control 
Unit PM1403 or PC 

RS-485 or USB interface 

Содержание PM1403

Страница 1: ...LPHA BETA RADIATION DETECTOR BDAB OPERATION MANUAL...

Страница 2: ...1 Switch on off BDAB 15 6 2 Operation modes 17 6 2 1 Measurement mode of radiation flux density Measurement Mode 18 6 2 2 Measurement mode of radiation flux density Measurement Mode 20 6 2 3 Search m...

Страница 3: ...nation as well as other necessary information for correct instrument operation and full utilization of its capabilities BDAB is documented as follows according to state standard of manufacturer countr...

Страница 4: ...iation DER search ionization sources to prevent the illicit trafficking of radioactive materials The instruments are also used by specialists in various industries including agriculture transportation...

Страница 5: ...2 1 5 m 1 1 Cable No 2 1 25 m 1 1 Cable No 2 2 0 25 m 1 1 Cable No 3 1 filter 10 Handle 1 1 Nozzle1 1 Certificate2 1 CD Software Operation Manual 1 1 Package 3 1 1 Supplied by user request by separat...

Страница 6: ...500 E max Figure 3 1 Typical dependence of BDAB sensitivity relative to radiation energy 3 6 Measurement range of radiation density flux from 1 0 to 5 105 min 1 cm 2 3 7 Permissible limits of main mea...

Страница 7: ...dditional relative error of radiation flux density measurement don t exceed 30 under exposure of external gamma radiation field of 200 R h exposure rate and don t exceed 15 under exposure of external...

Страница 8: ...o electrostatic discharges test level 3 air discharge of 8 kV voltage contact discharge of 6 kV voltage performance criterion is resistant to is resistant to electromagnetic interference caused by rad...

Страница 9: ...as a monoblock BDAB is enclosed in a shockproof container Detector of radiation is based on the proportional counter with a mica window General BDAB construction is presented in Figure 4 1 a b a fron...

Страница 10: ...radiation detector 10 protection screen for input window of radiation detector Screen protects against mechanical injuries Center of the sensitive surface of BDAB is marked with the X sign on the fro...

Страница 11: ...radiation background A built in detector based on the proportional counter with a mica window is used to register and radiation flux density DER count rate Built in BDAB microcontroller processes imp...

Страница 12: ...manufacturer s name manufacturer s trademark serial number according to the numbering system of the manufacturer year of manufacture IP protection level state register mark 4 5 Package BDAB is packed...

Страница 13: ...tect the detector s entrance window radiation from mechanical damage to prevent breakdown of the BDAB detector Figure 4 1 positions 6 10 Always protect BDAB against blows and mechanical injuries corro...

Страница 14: ...ion 4 Check ability of BDAB to switch to every operation mode ALPHA LED lights blue indicates that the measurement mode of radiation flux density is activated Figure 4 1 position 1 BETA LED lights yel...

Страница 15: ...R LED will light on the front panel of BDAB Figure 4 1 position 4 1 alarm sounder produces sound audio alarm when set DER DE thresholds are exceeded 2 ALARM LED for alarming 3 BAT LED for On Off state...

Страница 16: ...ighlighted option of the MENU 4 MODE BACK OK open the list of Spectroscopic Radiation Monitor 1403 operation modes Exit or return to the previous operation mode or option Directional pad keys 5 UP 6 D...

Страница 17: ...Monitor 1403 switches off 1 2 Figure 6 3 BDAB enables the following operation modes measurement of radiation flux density measurement of radiation flux density search for radiation sources program op...

Страница 18: ...ect BDAB to Spectroscopic Radiation Monitor 1403 and enter the indication mode of BDAB operation modes To do it select the Detector BDAB line and press SELECT or ENTER key on Spectroscopic Radiation M...

Страница 19: ...c Radiation Monitor 1403 When statistical error exceeds 10 the Spectroscopic Radiation Monitor 1403 S LCD will display a corresponding message Figure 6 6 4 If current measurement error is sufficient t...

Страница 20: ...tor BDAB line and press SELECT or ENTER key on Spectroscopic Radiation Monitor 1403 Operation sequence in this mode is as follows 1 select the measurement mode of radiation flux density Measurement Mo...

Страница 21: ...r installation 8 press OK 9 Spectroscopic Radiation Monitor 1403 s LCD will indicate continuously measured flux density values of radiation in min 1 cm2 Figure 6 7 5 User can read the measured value o...

Страница 22: ...sources Figure 6 8 1 or radiation Figure 6 8 2 BDAB with Spectroscopic Radiation Monitor 1403 can as well be used to search for radiation sources But this mode is recommended when the count rate excee...

Страница 23: ...witch to the search mode automatically set detection threshold Thresholds 6 3 BDAB settings Select Detector BDAB line Figure 6 9 1 in the Spectroscopic Radiation Monitor 1403 s menu Instrument Setting...

Страница 24: ...When Display and Control Unit 1403 is switched on BDAB is energized automatically At that the POWER LED will light on the front panel of BDAB Figure 4 1 position 4 1 color LCD 3 5 89 mm 2 instrument s...

Страница 25: ...l Unit 1403 to switch off 7 2 Operation modes BDAB enables the following operation modes measurement of radiation flux density measurement of radiation flux density search for radiation sources progra...

Страница 26: ...s cursor keys Use SELECT to switch to the operation mode 1 2 3 Figure 7 4 7 2 1 Measurement mode of radiation flux density Measurement Mode To measure radiation flux density connect BDAB to Display an...

Страница 27: ...displayed graphically on the analogue scale as well Quantity of the lighting scale segments corresponds to the value of measured radiation flux density relative to the set threshold level When set thr...

Страница 28: ...r is 10 or less Current value of count rate is compared to the operation threshold calculated in the calibration mode If current value of count rate exceeds the threshold value then light sound audio...

Страница 29: ...is given in Table 10 1 Table 10 1 Possible problem Possible reasons Troubleshooting Green ALFA or BETA LED doesn t light when BDAB is being connected to Spectroscopic Radiation Monitor PM1403 Display...