Polaroid ID3000 Service Manual
Calibration Procedures
4-17
5.
When the Working . . . message disappears and Enter Calibration Data is high
lighted, the chart is ready. From the Camera Back, pull the white tab, then the
yellow tab to start the film processing Timer.
NOTE: This is an illustrative example only, to show how tick marks are
counted. Your calibration print may have different tick mark
counts.
Nominal values of R, G and B are about 35 +_3.
6.
Next select Enter Calibration Data and follow the screen instructions for entering
the information from the calibration print just produced. Count the number of tick
marks along the red wedge, going from the thick, left end to the right end where it
tapers off the edge.
Type the number of tick marks next to RED WEDGE and press <Enter>. The
highlight will automatically move to the next color. Repeat the process for each
color wedge, pressing <Enter> after each number entered. When all data has
been entered, press <F10>.
Содержание ID-3000
Страница 2: ...This page intentionally blank ...
Страница 4: ...This page intentionally blank ...
Страница 6: ...Polaroid ID3000 Service Manual Description 1 2 This page intentionally blank ...
Страница 24: ...Polaroid ID3000 Service Manual System Installation 2 2 This page intentionally blank ...
Страница 32: ...Polaroid ID3000 Service Manual System Installation 2 10 Figure 2 7 ID 3000F System Cable Connections ...
Страница 33: ...Polaroid ID3000 Service Manual System Installation 2 11 Figure 2 8 ID 3000T System Cable Connections ...
Страница 84: ...This page intentionally blank ...
Страница 130: ...This page intentionally blank ...
Страница 174: ...This page intentionally blank ...
Страница 198: ...This page intentionally blank ...
Страница 200: ...This page intentionally blank ...
Страница 254: ...This page intentionally blank ...
Страница 272: ...This page intentionally blank ...