background image

 Warranty card 

Product Name 

Portable Nebulizer 

Model Number 

      

Serial Number
Purchase Date

 Dealer

1.  Please get a shop stamp from your dealer, put the purchase date on this card, and keep this card with 

you. Presentation of this card is required when a warranty service is requested. (No new card will be 

issued to you if this card is lost.)

2.  Microbase guarantees that the main unit of Pocket Air, if used according to the instructions, will be 

free from defects in material and workmanship caused by the manufacturing process for the warranty 

period of 2 years. Any other replacement parts or accessories are not covered hereby. You will void 

this warranty if you disassemble the main unit of this Portable Nebulizer.

3.  The warranty period of the medication cup will be 12 months, starting from the purchase date. The 

warranty does not apply to ( what’s NOT covered) : 

I . Mesh clogging. 

II . Any failure to follow our written instructions for the product. 

III . Damage or defect due to willful neglect or negligence by anyone other than us.

4.  During the warranty period, Microbase will repair or replace the device if a defect is discovered. If it is 

replaced, the replacement device may either be the same model or a model that is at least comparably 

equipped. Replacement or repair of the device shall not serve as the basis for a new warranty. All 

replaced old devices or parts shall become the property of Microbase. Further claims are excluded. 

This disclaimer of warranty shall be ineffective in the event of injury to life, limb or health, in cases of 

wrongdoing and gross negligence, product liability and if substantive obligations under the warranty 

agreement are violated. Any expenses of non-maintenance work (including but not limited to, the back 

and forth shipping cost or expenses arising from your specific request) will be borne by you.

5.  The warranty shall be cancelled if:

-the device has been operated or used improperly with respect to the descriptions in the 

instructions for use

-damage is present that is attributable to the effects of water, fire, lightning, etc.

-the damage was caused by transporting the device incorrectly or a falling impact

-the device has been misused or not cared for correctly

-repairs, adaptations or modifications have been made to the device by persons not authorized 

by Microbase

6.  For any service  requested without this warranty card, beyond the warranty period, or not covered by 

this warranty, a service fee will be charged.

7.  In the event that replacement parts or warranty service is needed, please contact the nearest dealer or 

the store where you purchased this device. 

HT12-17008-01

Содержание MB0500300

Страница 1: ...MBPN002 MB0500300 MB05006 Instruction Manual Portable Nebulizer POCKET AIR R ...

Страница 2: ...nts 5 Optional Parts and Accessories 5 System Overview 6 How to set up the Medication cup 7 Connecting to power supply Applicable to MBPN002 MB0500300 8 Connecting to power supply Applicable to MB0500600 10 How to inhale properly 12 Cleaning and Disinfection 13 Troubleshooting 15 Specifications 16 Technical Data 18 ...

Страница 3: ...l can be operated with power adapter or with batteries or rechargeable batteries Pocket Air does not charge rechargeable batteries rechargeable batteries should be charged separately Portable Nebulizer Model No MB0500300 Portable Nebulizer Model No MBPN002 MB05006 Please read the instructions carefully and completely Failure to comply with instructions for use may result in damage to the device In...

Страница 4: ...for patients who are unconscious not breathing spontaneously or have pulmonary edema Babies and children must be supervised by an adult during treatment Recommended operation environment This device is intended for use in medical facility such as hospital clinic and doctor s office and in general household Temperature range 10 C to 40 C Humidity 30 to 85 RH Precautions for use Warnings and caution...

Страница 5: ...dividuals who require special assistance For safety reasons always disconnect the power adapter from the socket under the following circumstances if a malfunction occurs during operation before cleaning the device immediately after use The product must not be operated and the power adapter must be unplugged from the socket immediately if the power adapter is damaged or if a fault is suspected Be s...

Страница 6: ... source Direct Current Operating temperature limits 10 C to 40 C Storage and delivery temperature limits 20 C to 70 C Operating humidity limits 30 to 85 R H Storage and delivery humidity limits 20 to 75 R H Manufacturer Date of manufacture Serial number Batch code Disposal of Electrical Electronic Equipment WEEE Do not treat this product as household waste This device compiles with the requirement...

Страница 7: ...ual Optional Parts and Accessories Medication Cup MBPN00201 MB0500601 Main Unit MBPN00200 MB0500300 MB0500600 Mask Child MBPN00204 Mask Adult MBPN00203 Power Adapter MBPN00205 Mouthpiece MBPN00207 USB Cable MBPN00208 Single patient use Single patient use Single patient use Available on MBPN002 MB05006 Available on MBPN002 MB05006 To order replacement parts or accessories please contact your local ...

Страница 8: ... Battery Cover Main Unit Power Cable Jack Power Cable Jack Cover Bottom View Light Indications Press ON OFF Button to turn on the power Press Release Button to remove the medication cup Green light on Ongoing operation Orange light on flash Low power Malfunction No light Malfunction Bottom cover Model no MB0500300 Model no MBPN002 Model no MB05006 ...

Страница 9: ...n cup Check to ensure that the buckle is aligned and inserted correctly Make sure that the indicator light is off before adding the medication Open the medication cup cap Add the quantity of medication prescribed by your doctor DO NOT EXCEED THE LIMIT Max 6ml Close the medication cup cap properly X Clean all parts of your portable nebulizer before use after each use and after extended storage ...

Страница 10: ...ondary lithium batteries shall comply with the requirements of IEC 62133 1 Do NOT insert a battery with the positive and negative poles reversed 2 Do NOT use old and new batteries together or mix and use different brand of batteries together Battery life and battery replacement 1 Nebulizer can be operated with alkaline batteries or nickel metal hydride NiMH rechargeable batteries 2 When the power ...

Страница 11: ... main unit Plug the adapter into a standard power socket The power adapter is an optional accessory and the appearance might be different Insert the jack connector of the USB cable into the cable jack on the main unit Plug the USB cable into a standard charger The standard charger is an optional accessory and the appearance might be different DC 5V 1A DC 5V 1A 1 2 1 The connection jack should be 9...

Страница 12: ...two AA nickel metal hydride NiMH rechargeable batteries Secondary lithium batteries shall comply with the requirements of IEC 62133 1 Do NOT insert a battery with the positive and negative poles reversed 2 Do NOT use old and new batteries together or mix and use different brands of batteries together 3 Do NOT pull the bottom cover too hard Battery life and battery replacement 1 Nebulizer can be op...

Страница 13: ...r into a standard power socket The power adapter is an optional accessory and the appearance might be different Open the bottom cover Insert the jack connector of the USB cable into the cable jack on the main unit Plug the USB cable into a standard charger The standard charger is an optional accessory and the appearance might be different DC 5V 1A DC 5V 1A 1 The connection jack should be 9 5mm in ...

Страница 14: ...art nebulization light Indicator turns green Place the mouthpiece between your teeth with your lips firmly sealed around the mouthpiece Accessories such as additional mouthpiece and mask could help to provide a more comfortable inhalation treatment If mask is used gently press the mask against the face so that it fits snugly over the mouth and nose Breathe in and out slowly through your mouth unti...

Страница 15: ...se the medication cup with distilled water 7 Shake off excess water and allow parts to be fully air dried on a clean dry towel 8 Use gauze or clean towel to wipe off stains on the main unit if necessary 9 Make sure that all cleaned parts are completely dry before you store or use them next time 1 Hand wash the mask and the mouthpiece in warm 30 C water using neutral detergent and gently rub it for...

Страница 16: ...for a prolonged period Failure to do so may result in damage due to battery leakage 3 Do NOT leave the device under direct sunlight or in high humidity extreme heat or cold environment 4 Keep this device away from fire high electromagnetic fields and out of reach of children Storage Medication cup Main unit Power adapter USB cable The following item can be disinfected by boiled water or alcohol Pa...

Страница 17: ...nstant orange light Low battery power Insert new batteries or connect the power adapter Power indicator shows flashing orange light Faulty connection between the electrodes and the main unit Use rubbing alcohol to clean the electrodes and reinstall the medication cup The mesh is damaged Replace the medication cup The medication cup is not installed properly Refer to How to Set up the medication cu...

Страница 18: ...roduct Portable Nebulizer Model MBPN002 MB0500300 Method of operation Active Vibrating Mesh Technology Power supply AA Alkaline Battery x 2 AA NiMH rechargeable Battery x 2 Power Adapter Power consumption 1 5W MBPN00201 2 4W MB0500601 Vibrating frequency Approx 117 kHz 15 Nebulization rate 0 25 ml min Particle size MMAD 5um Capacity of medication cup Approx 6 ml Dimension Approx L78 X W41 X H73 mm...

Страница 19: ...ications Product Portable Nebulizer Model MB05006 Method of operation Active Vibrating Mesh Technology Power supply AA Alkaline Battery x 2 AA NiMH rechargeable Battery x 2 Power Adapter Power consumption 1 5W MBPN00201 2 4W MB0500601 Vibrating frequency Approx 117 kHz 15 Nebulization rate 0 25 ml min Particle size MMAD 5um Capacity of medication cup Approx 6 ml Dimension Approx L81 X W44 X H77 mm...

Страница 20: ...therapy equipments Part 1 Nebulizing systems and their components Annex CC 3 using multistage cascade impactor to measure particle size Model MBPN00201 Particle Size MMAD 5um MMAD Mass Median Aerodynamic Diameter Medication Cup Capacity 6 ml max Noise Less than 50dB Around 1 meter ...

Страница 21: ...19 Technical Data Model MB0500601 Particle Size MMAD 5um MMAD Mass Median Aerodynamic Diameter Medication Cup Capacity 6 ml max Noise Less than 50dB Around 1 meter ...

Страница 22: ...the electromagnetic environment specified below The customer or the user of the MBPN002 MB0500300 should assure that it is used in such an environment Emission test Compliance Electromagnetic environment guidance RF emissions CISPR 11 Group 1 The MBPN002 MB0500300 uses RF energy only for its internal function Therefore its RF emissions are very low and are not likely to cause any interference in n...

Страница 23: ...e s to earth 1kV differential mode Not applicable Mains power quality should be that of a typical commercial or hospital environment Voltage Dips short interruptions and voltage variations on power supply input lines IEC 61000 4 11 5 UT 95 dip in UT for 0 5 cycle 40 UT 60 dip in UT for 5 cycles 70 UT 30 dip in UT for 25 cycles 5 UT 95 dip in UT for 5 sec 5 UT 95 dip in UT for 0 5 cycle 40 UT 60 di...

Страница 24: ...xed RF transmitters as determined by an electromagnetic site surveya should be less than the compliance level in each frequency rangeb Interference may occur in the vicinity of equipment marked with the following symbol NOTE1 At 80 MHz and 800 MHz the higher frequency range applies NOTE2 These guidelines may not apply in all situations Electromagnetic propagation is affected by absorption and refl...

Страница 25: ... of transmitter W Separation distance according to frequency of transmitter m 150 kHz to 80 MHz d 1 2 P 80 MHz to 800 MHz d 1 2 P 800 MHz to 2 5 GHz d 2 3 P 0 01 0 12 0 12 0 23 0 1 0 38 0 38 0 73 1 1 2 1 2 2 3 10 3 8 3 8 7 3 100 12 12 23 For transmitters rated at a maximum output power not listed above the recommended separation distance d in metres m can be estimated using the equation applicable...

Страница 26: ...ass B The MB05006 is suitable for use in all establishments including domestic establishments and those directly connected to the public low voltage power supply network that supplies buildings used for domestic purposes Harmonic emissions IEC 61000 3 2 Class A Voltage fluctuations flicker emissions IEC 61000 3 3 Compliance In order to regulate the requirements for EMC with the aim to prevent unsa...

Страница 27: ...ge IEC 61000 4 5 0 5kV 1kV line s to line s 0 5kV 1kV 2kV line s to earth 0 5kV 1kV line s to line s Not applicable Mains power quality should be that of a typical home healthcare environment Voltage Dips short interruptions and voltage variations on power supply input lines IEC 61000 4 11 Voltage dips 0 U T 0 5 cycle 0 U T 1 cycle 70 U T 25 30 cycles Voltage interruptions 0 U T 250 300 cycle Volt...

Страница 28: ... watts W according to the transmitter manufacturer and d is the recommended separation distance in metres m Field strengths from fixed RF transmitters as determined by an electromagnetic site survey a should be less than the compliance level in each frequency range b Interference may occur in the vicinity of equipment marked with the following symbol NOTE1 At 80 MHz and 800 MHz the higher frequenc...

Страница 29: ...0 810 800 960 GSM 800 900 TETRA 800 iDEN 820 CDMA 850 LTE Band 5 Pulse modulation b 18 Hz 2 0 3 28 28 870 930 1720 1700 1990 GSM 1800 CDMA 1900 GSM 1900 DECT LTE Band 1 3 4 25 UMTS Pulse modulation b 217 Hz 2 0 3 28 28 1845 1970 2450 2400 2570 Bluetooth WLAN 802 11 b g n RFID 2450 LTE Band 7 Pulse modulation b 217 Hz 2 0 3 28 28 5240 5100 5800 WLAN 802 11 a n Pulse modulation b 217 Hz 0 2 0 3 9 9 ...

Страница 30: ...quation applicable to the frequency of the transmitter where p is the maximum output power rating of the transmitter in watts W according to the transmitter manufacturer NOTE1 At 80 MHz and 800 MHz the separation distance for the higher frequency range applies NOTE2 These guidelines may not apply in all situations Electromagnetic propagation is affected by absorption and reflection from structures...

Страница 31: ...that is at least comparably equipped Replacement or repair of the device shall not serve as the basis for a new warranty All replaced old devices or parts shall become the property of Microbase Further claims are excluded This disclaimer of warranty shall be ineffective in the event of injury to life limb or health in cases of wrongdoing and gross negligence product liability and if substantive ob...

Страница 32: ......

Отзывы: