Recommended separation distance between
portable and mobile RF communications equipment and the MB05006
The MB05006 is intended for use in an electromagnetic environment (for home healthcare) in which
radiated RF disturbances are controlled. The customer or the user of the MB05006 can help prevent
electromagnetic interference by maintaining a minimum distance between portable and mobile RF
communications equipment (transmitters) and the MB05006 as recommended below, according to the
maximum output power of the communications equipment.
Rated maximum output
power of transmitter
W
Separation distance according to frequency of transmitter
m
150 kHz to 80 MHz
d =1,2
√
P
80 MHz to 800 MHz
d =1,2
√
P
800 MHz to 2,5 GHz
d =2,3
√
P
0,01
N/A
0,12
0,23
0,1
N/A
0,38
0,73
1
N/A
1,2
2,3
10
N/A
3,8
7,3
100
N/A
12
23
For transmitters rated at a maximum output power not listed above, the recommended separation distance
d in meters (m) can be estimated using the equation applicable to the frequency of the transmitter,
where p is the maximum output power rating of the transmitter in watts (W) according to the transmitter
manufacturer.
NOTE1 At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
NOTE2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by
absorption and reflection from structures, objects and people.
WEEE (Directive on Wasted Electrical and Electronic Equipment)
This marking shown on the product indicates that it should not be disposed of, with other household wastes at
the end of its working life. To prevent possible harm to the environment or human health from uncontrolled waste
disposal, please separate this from other types of wastes and recycle it responsibly to promote the sustainable reuse
of material resources.
Household users should contact either the retailer where they purchased this product, or their local government
office, for details of where and how they can take this equipment for safe recycling.