25
Technical Data
Manufacturer’s declaration-electromagnetic immunity
The MB05006 is intended for use in the electromagnetic environment (for home healthcare) specified below.
The customer or the user of the MB05006 should assure that it is used in such an environment.
Immunity test
IEC 60601 test level
Compliance level
Electromagnetic
environment-guidance
(for home healthcare environment)
Electrostatic
discharge(ESD) IEC
61000-4-2
Contact:±8 kV
Air±2 kV,±4 kV,±8
kV,±15 kV
Contact:±8 kV
Air±2 kV,±4 kV,±8
kV,±15 kV
Floors should be wood,
concrete or ceramic tile.
If floors are covered with
synthetic material, the relative
humidity should be at least
30%.
Electrical fast
transient/burst IEC
61000-4-4
+ 2kV for power
supply lines
+ 1kV for input/output
lines
+ 2kV for power
supply lines
Not applicable
Mains power quality should
be that of a typical home
healthcare environment.
Surge IEC 61000-4-5
+ 0.5kV, +1kV line(s)
to line(s)
+ 0.5kV, +1kV,+ 2kV
line(s) to earth
+ 0.5kV, +1kV line(s)
to line(s)
Not applicable
Mains power quality should
be that of a typical home
healthcare environment.
Voltage Dips, short
interruptions and
voltage variations on
power supply input
lines IEC 61000-4-11
Voltage dips:
0 % U
T
; 0,5 cycle
0 % U
T
; 1 cycle
70 % U
T
; 25/30
cycles
Voltage interruptions:
0 % U
T
; 250/300
cycle
Voltage dips:
0 % U
T
; 0,5 cycle
0 % U
T
; 1 cycle
70 % U
T
; 25/30
cycles
Voltage interruptions:
0 % U
T
; 250/300
cycle
Mains power quality should
be that of a typical home
healthcare environment. If the
user of the MB05006 requires
continued operation during
power mains interruptions,
it is recommended that the
MB05006 be powered from an
uninterruptible power supply or
a battery.
Power frequency(50,
60 Hz) magnetic field
IEC 61000-4-8
30 A/m
50 Hz or 60 Hz
30 A/m
50 Hz
The MB05006 power frequency
magnetic fields should be at
levels characteristic of a typical
location in a typical home
healthcare environment.
NOTE UT is the a.c. mains voltage prior to application of the test level.