Technical Data
Manufacturer’s declaration-electromagnetic immunity
The MB05006 is intended for use in the electromagnetic environment (for home healthcare) specified below.
The customer or the user of the MB05006 should assure that is used in such and environment.
Immunity test IEC 60601 test level Compliance level
Electromagnetic environment-guidance (for home
healthcare environment)
Conducted RF
IEC 61000-4-6
Radiated RF
IEC 61000-4-3
3 Vrms:
0,15 MHz – 80 MHz
6 Vrms:
in ISM and amateur
radio bands between
0,15 MHz and 80 MHz
80 % AM at 1 kHz e)
10 V/m
80 MHz – 2,7 GHz b)
80 % AM at 1 kHz c)
3 Vrms:
0,15 MHz – 80 MHz
6 Vrms:
in ISM and amateur
radio bands between
0,15 MHz and 80
MHz
80 % AM at 1 kHz e)
10 V/m
80 MHz – 2,7 GHz
80 % AM at 1 kHz
Portable and mobile RF communications equipment should be
used no closer to any part of the MB05006 including cables,
than the recommended separation distance calculated from the
equation applicable to the frequency of the transmitter.
Recommended separation distance:
d = 1,2 √P
d = 1,2 √P 80MHz to 800 MHz
d = 2,3 √P 800MHz to 2,7 GHz
Where P is the maximum output power rating of the transmitter in
watts (W) according to the transmitter manufacturer and d is the
recommended separation distance in metres (m).
Field strengths from fixed RF transmitters, as determined by an
electromagnetic site survey,
a
should be less than the compliance
level in each frequency range.
b
Interference may occur in the vicinity of equipment marked with
the following symbol:
NOTE1
At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from
structures, objects and people.
a. Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile radios,
amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy. To assess the
electromagnetic environment due to fixed RF transmitters, an electromagnetic site survey should be considered. If the measured
field strength in the location in which the MB05006 is used exceeds the applicable RF compliance level above, the MB05006
should be observed to verify normal operation. If abnormal performance is observed, additional measures may be necessary, such
as re-orienting or relocating the MB05006.
b. Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m.