Page 22
Practical Instrument Electronics
Tel: 585.872.9350 • Fax: 585.872.2638 • [email protected] • www.piecal.com
Operating Instructions
Source Thermocouple
Choose this function to provide a simulated thermocouple signal into controllers,
temperature transmitters, indicators or any input devices that measure
thermocouple sensors.
Move the power switch
w
to SOURCE then Double Click the EZ-DIAL knob to
get into the Double Click Menu. Turn the knob to scroll through the settings and
press the knob to make your selection. Select T/C for the FUNCTION, °F or °C
for the UNITS, T/C Type (J, K, T, E, R, S, B, N, G, C, D, L (J-DIN), U (T-DIN) or
P (Platinel II) and internal COLD JUNC ON or OFF (ON is the default).
Connect the 830PM to the inputs of the device being calibrated using the proper
type of thermocouple wire via the miniature thermocouple socket.
Instantly output your SPAN and ZERO output settings by moving the EZ-CHECK
switch between HI and LO. You may also select any third output setting (such as
mid-range) using the SET position on the EZ-CHECK switch. The output is
adjusted in 0.1° increments by turning the knob
e
. Press and turn the knob for
faster dialing with 10.0° increments.
5
MAX
READ
MIN
READ
6
-
+
-
3 2
-2
OFF
HI
SET
LO
SOURCE
Hz SYNC
EZ-DIAL
Double Click
Menu
Push & Turn for
Fast Dialing
Push & Hold to
Store/Step
piecal.com
EZ-CHECK™
PRACTICAL INSTRUMENT ELECTRONICS
MODEL 830PM
Process Calibrator
Loop Diagnostics · Transmitter Supply
mA · V ∙ pH · TC · Ω · RTD · Hz · Pressure
Pressure
Module
All mA∙ Read V
TC
Source & Read mV
Hz
Source V
pH
RTD • Ohms
OVERLOAD
4
1
+
+1
Instrument with T/C Input
Controller
Temperature Transmitter
Temperature Indicator
Temperature Trip or Alarm
Thermocouple Wire
T/C
+
-
TYPE K °C
830.0
LO