![PIE PIE 830PM Скачать руководство пользователя страница 25](http://html1.mh-extra.com/html/pie/pie-830pm/pie-830pm_manual_1555584025.webp)
Page 23
Practical Instrument Electronics
Tel: 585.872.9350 • Fax: 585.872.2638 • [email protected] • www.piecal.com
Operating Instructions
Read Thermocouple Sensors
Choose this function to measure temperatures with a thermocouple probe or
sensor.
Move the power switch
w
to READ then Double Click the EZ-DIAL knob to get
into the Double Click Menu. Turn the knob to scroll through the settings and
press the knob to make your selection. Select T/C for the FUNCTION, °F or °C
for the UNITS, T/C Type (J, K, T, E, R, S, B, N, G, C, D, L (J-DIN), U (T-DIN) or
P (Platinel II) and COLD JUNC ON or OFF (ON is the default).
Connect the 830PM to the inputs of the device being calibrated using the proper
type of thermocouple wire via the miniature thermocouple socket. If no sensor is
connected, a wire is broken or the sensor is burned out, OPEN TC will appear
on the display. Signals above the maximum scale are limited by protection
circuitry with “OVER RANGE” on the display and the red OVERLOAD LED lit.
The 830PM measures the input signal and constantly updates the display with
the current reading. Move the EZ-CHECK switch
q
to MAX to see the highest
reading and to MIN to see the lowest reading. Press and hold the knob
e
to clear
the MAX and MIN readings.
5
MAX
READ
MIN
READ
6
-
+
-
3 2
-2
OFF
HI
SET
LO
SOURCE
Hz SYNC
EZ-DIAL
Double Click
Menu
Push & Turn for
Fast Dialing
Push & Hold to
Store/Step
piecal.com
EZ-CHECK™
PRACTICAL INSTRUMENT ELECTRONICS
MODEL 830PM
Process Calibrator
Loop Diagnostics · Transmitter Supply
mA · V ∙ pH · TC · Ω · RTD · Hz · Pressure
Pressure
Module
All mA∙ Read V
TC
Source & Read mV
Hz
Source V
pH
RTD • Ohms
OVERLOAD
4
1
+
+1
Thermocouple Wire
T/C
+
-
TYPE K °F
1830.1