Page 31
Practical Instrument Electronics
Tel: 585.872.9350 • Fax: 585.872.2638 • [email protected] • www.piecal.com
Operating Instructions
Read Pressure
Media Compatibility
Non-isolated DN sensors: clean, dry, non-corrosive, non-condensing gases only
Isolated DI sensors: any media compatible with 316L SS & Viton®
Isolated GI, CI & AI sensors: any media compatible with 316L SS
Pressure Module Accuracy
±0.025% of full scale including all effects of linearity, repeatability and hysteresis
from -20° to +50°C (-4° to +122°F)
* The DN0010 sensor accuracy is ±0.050% of full scale
32 Engineering Units:
PSI • inches, feet, mm, cm and meter of H2O @ 4°C, 20°C & 60°F inches,
meter, cm and mm of Hg @ 0°C
torr • kg/cm2 • kg/m2 • Pa • hPa • kPa • MPa • Bar • mBar • ATM • oz/in2 • lb/ft2
Sensor Code
Application
DNxxxx
Differential, Non-isolated
0 to 0010*, 0028, 0200, 0415, 2000” H2O
DIxxxx
Differential, Isolated
0 to 0001, 0005, 0015, 0030, 0100, 0300, 0500 PSID
GIxxxx
Gauge, Isolated
0 to 0015, 0030, 0050, 0100, 0300, 0500, 1000, 3000 PSIG
CIxxxx
Compound, Isolated
-14.7 to +0015, 0030, 0050, 0100, 0300, 0500, 1000, 3000 PSIG
AIxxxx
Absolute, Isolated
0 to 0017, 0038, 0100, 1000 PSIA
Optional Pressure Modules