5 Installation
V-273 VC Linear Actuator
MP155E
Version: 1.1.4
17
In this Chapter
General Notes on Installation ...................................................................................................... 17
Installing the V-273 into a Mechanical Mounting ....................................................................... 20
Connecting the V-273 to the Controller ...................................................................................... 23
5.1
General Notes on Installation
DANGER
Strong magnetic fields affect heart pacemakers!
The V-273 contains permanent magnets that could impair the function of heart pacemakers
and electronic implants.
The permanent magnet's magnetic fields are also effective when the
linear actuator is not connected or switched off.
Make sure that people with heart pacemakers and electronic implants do not have access
to the V-273.
Make sure that the danger zone is marked.
CAUTION
Risk of cuts and crushing!
The V-273 linear actuator can generate high forces at high velocities. When a contact part with
a small contact surface is installed on the runner of the linear actuator, the pressure exerted by
the runner on a fixed part or obstacle during operation may be very high. If fingers or limbs get
caught between the runner of the linear actuator and a fixed part or obstacle during operation,
there is a risk of minor injury from cuts and crushing.
Use protective structures to keep limbs away from areas in which they could be caught by
moving parts.
Pay attention to the safety distances in accordance with the applicable standards when
installing protective structures.
If possible, use contact parts with a sufficiently dimensioned contact area, so that a
pressure of 50 N / cm
2
is not exceeded when operating the linear actuator with the
maximum specified force (p. 37).
5
Installation