Confirm cardiac arrest and start manual CPR with a minimum of interruptions until
LUCAS is applied and ready.
Always follow local and/or international guidelines for CPR when you use LUCAS.
ON/OFF
ADJUST
PAUSE
ACTIVE
(continuous)
ACTIVE
(30:2)
QUICK REFERENCE GUIDE
LUCAS is manufactured by JOLIFE AB in Sweden and distributed worldwide by Physio-Control, Inc.
For further information, please visit our website at www.physio-control.com.
NOTE:
This guide is not a complete Instructions for Use. Refer to the “Instructions for Use” for complete
directions for use, indications, contraindications, warnings, precautions, and potential adverse events.
LUCAS™ 2 Chest Compression System –
Instructions for Use
100901-00 Rev A, Valid from CO J2716 © 2014 Jolife AB
Activate (A)
s
Push
ON/OFF
for 1 second to
start self-test and power up
LUCAS
Back Plate (B)
s
Pause manual CPR
s
Carefully put Back Plate under
the patient, below armpits
s
Resume manual CPR
Compressor (C)
s
Pull release rings once; claw
locks open. Then let go of the
release
rings
s
Attach to Back Plate; listen
for
“click”
s
Pull up once to ensure
attachment
Position the
Suction Cup
s
Center the Suction Cup over
the
chest
s
The lower edge of Suction Cup
should be immediately above
the end of the sternum
Push down the
Suction Cup
s
Push the Suction Cup down
with two fingers (make sure
it is in the
ADJUST
mode)
s
Pressure pad inside Suction
Cup should touch patient’s
chest. If the pad does not
touch or fit properly, continue
manual
compressions
s
Push
PAUSE
to lock Start
Position – then remove your
fingers from the Suction Cup
Start compressions
s
Check for proper position.
Adjust
if
necessary
s
Push
ACTIVE (continuous)
or
ACTIVE (30:2)
s
LUCAS provides chest
compressions according to
guidelines.
Attach the
Stabilization Strap
s
Attach the LUCAS Stabilization
Strap
100901-00 Rev B, Valid from CO J3166 © 2017 Jolife AB