LUCAS
TM
2 Chest Compression System –
Instructions for Use
100901-00 Rev A, Valid from CO J2716 © 2014 Jolife AB
30
9.5 Battery physical specifications
9.6 Battery environmental specifications
9.7 Power supply specification (optional accessory)
Model - MVB100024A
Model - MWA150028B
Category
Specifications
Size (H × W × D)
5.1 x 3.5 x 2.2 inches / 13.0 × 8.8 × 5.7 cm
Weight
1.3 lbs / 0.6 kg
Type
Rechargeable Lithium Ion Polymer (LiPo)
Capacity
3300 mAh (typical), 86 Wh
Battery voltage (nominal)
25.9 V
Initial Battery runtime (nominal patient)
45 minutes (typical)
Maximum Battery charge time
Less than 4 hours at room temperature
(72ºF / 22ºC)
Required interval for replacement of the Battery
Recommendation to replace the Battery every 3 years or
after 200 uses (of more than 10 minutes use each time)
Category
Specifications
Operating temperature
32ºF to 104ºF / 0ºC to +40ºC, ambient when installed in the
device
Charge temperature
41ºF to 95ºF / 5ºC to +35ºC ambient (68ºF to 77ºF / 20ºC to
25ºC preferred)
Storage temperature
32ºF to 104ºF / 0ºC to +40ºC ambient for less than six
months
IP classification (IEC60529)
IP 44
Category
Specifications
Input
100-240VAC, 50/60Hz, 2.3A
Output 24VDC,
4.2A
Category
Specifications
Input
100-240 VAC, 50/60Hz, 2A
Output 28VDC,
5.4A