8.3 Malfunction alarms
Below is a list of all alarms that can occur on LUCAS. You mute all alarms for 60 seconds if you
push
MUTE
.
Start with manual compressions immediately if LUCAS does not operate properly.
If the malfunction described above seems permanent, LUCAS must be examined by approved
service personnel. Please consult your local Physio-Control representative. Contact information
is available at www.physio-control.com.
Priority Reason
Visual LED indication
Audible alarms
Result
N/A
Rising temperature in
LUCAS
None
Information Signal
None
High
Priority
Compression pattern
outside limit (too deep,
too shallow or timing
failure)
Intermittent red alarm
LED
High Priority Alarm
LATCHING ALARM SIGNAL
Compressions
stop
High
Priority
Too high temperature
in LUCAS
Intermittent red alarm
LED
High Priority Alarm
LATCHING ALARM SIGNAL
Compressions
stop
High
Priority
Hardware error
Intermittent red alarm
LED
High Priority Alarm
LATCHING ALARM SIGNAL
Compressions
stop
High
Priority
Too high Battery
temperature
Intermittent red alarm
LED
Red Battery alarm:
The red Battery
Indication LED
farthest to the right lit
intermittently.
High Priority Alarm
LATCHING ALARM SIGNAL
Compressions
stop
High
Priority
Battery charge too low
Intermittent red alarm
LED
Red Battery alarm:
The red Battery
Indication LED
farthest to the right lit
intermittently.
High Priority Alarm
LATCHING ALARM SIGNAL
Compressions
stop.
The Battery must
be recharged.
(4s)
(4s) ...
(..)
(...)
(..)
(..)
(..)
(2.5s)
(...)
(2.5s)
(..)
(...)
(..)
(..)
(..)
(2.5s)
(...)
(2.5s)
(..)
(...)
(..)
(..)
(..)
(2.5s)
(...)
(2.5s)
(..)
(...)
(..)
(..)
(..)
(2.5s)
(...)
(2.5s)
(..)
(...)
(..)
(..)
(..)
(2.5s)
(...)
(2.5s)
100901-00 Rev B, Valid from CO J3166 © 2017 Jolife AB