9.9 Electromagnetic environmental declaration
Guidance and manufacturer's declaration - electromagnetic emissions
LUCAS 2 is intended for use in the electromagnetic environment specified below. The customer or the operator of
LUCAS 2 must make sure that it is used in the correct environment.
Emissions test
Compliance
Electromagnetic
environment - guidance
RF emissions CISPR 11
Group 1
LUCAS 2 uses RF energy only for its
internal operation. This makes its RF
emissions very low and not likely to
cause interference with other electronic
equipment near LUCAS 2.
RF emissions CISPR 11
Class B
LUCAS 2 is suitable for use in all
buildings including domestic homes
and places directly connected to the
public low-voltage Power Supply
network that supplies buildings used
for domestic purposes.
Harmonic emissions IEC 61000-3-2
Class A
Voltage fluctuations/ flicker emissions
IEC 61000-3-3
Complies
Guidance and manufacturer's declaration - electromagnetic immunity
LUCAS 2 is intended for use in the electromagnetic environment specified below. The customer or the operator of
LUCAS 2 must make sure that it is used in the correct environment.
Immunity test
IEC 60601 test level
Compliance level
Electromagnetic
environment - guidance
Electrostatic
discharge (ESD)
IEC 61000-4-2
+/- 6 kV contact
+/- 8 kV air
+/- 6 kV contact
+/- 8 kV air
Floors must be wood, concrete or
ceramic tile. If there is synthetic
material on the floor, the relative
humidity must be 30% or more.
Electrical fast
transient / Burst
IEC 61000-4-4
+/- 2 kV for Power
Supply lines
+/- 1 kV for input/output lines
+/- 2 kV for Power
Supply lines
n/a. for input/output lines
The mains power quality must be
that of a typical commercial or
hospital environment.
Surge
IEC 61000-4-5
+/- 1 kV differential mode
+/- 2 kV common mode
+/- 1 kV differential mode
n/a. for common mode
The mains power quality must be
that of a typical commercial or
hospital environment.
Voltage dips, short in-
terruptions and voltage
variations on Power
Supply input lines
IEC 61000-4-11
<5 % UT (>95 % dip in UT)
for 0.5 cycle
40 % UT (60 % dip in UT)
for 5 cycles
70 % UT (30 % dip in UT)
for 25 cycles
<5 % UT (>95 % dip in UT))
for 5 sec
<5 % UT (>95 % dip in UT)
for 0.5 cycle
40 % UT (60 % dip in UT)
for 5 cycles
70 % UT (30 % dip in UT)
for 25 cycles
<5 % UT (>95 % dip in UT))
for 5 sec
The mains power quality must be
that of a typical commercial or
hospital environment. If the user of
the [Equipment or System] requires
continued operation during power
mains interruptions, JOLIFE recom-
mends that the [Equipment or
System] is energized from a Power
Supply or Battery that cannot be
interrupted.
Power frequency
(50/60 Hz) magnetic
field IEC 61000-4-8
30 A/m
30 A/m
The power frequency magnetic
fields must be at levels that are
characteristic of a typical location in
a typical commercial or hospital
environment
NOTE UT is the a.c. mains voltage prior to application of the test level.
Guidance and manufacturer's declaration - electromagnetic immunity
LUCAS 2 is intended for use in the electromagnetic environment specified below. The customer or the operator of LUCAS
2 must make sure that it is used in the correct environment.
100901-00 Rev B, Valid from CO J3166 © 2017 Jolife AB