LUCAS
TM
2 Chest Compression System –
Instructions for Use
100901-00 Rev A, Valid from CO J2716 © 2014 Jolife AB
22
5.7.2 Connect to the external Power
Supply
You can connect the LUCAS Power Supply
or Car Power Cable in all LUCAS operating
modes.
Caution - keep Battery installed
The Battery must always be installed for
LUCAS to be able to operate, also when
powered by the external Power Supply.
To use the Power Supply cable:
• Connect the Power Supply cable to
LUCAS.
• Connect the mains cable to the wall mains
outlet (100-240V, 50/60Hz)
To use the Car Power Cable:
• Connect the Car Power Cable to LUCAS
• Connect the Car Power Cable to the car
outlet (12-28VDC)
5.8 Adjunctive therapies
Caution - adjunctive therapies
The use of other medical equipment or drugs
in conjunction with LUCAS can affect the
treatment. Always consult the instructions for
use for the other equipment and/or drugs to
make sure that they are applicable in
conjunction with CPR.
5.8.1 Defibrillation
Defibrillation can be performed while LUCAS
operates.
1. You can apply the defibrillation elec-
trodes before or after LUCAS has been
put in position.
2. Perform the defibrillation according to
the instructions from the manufacturer of
the defibrillator.
Caution - defibrillation electrodes
Position the defibrillation electrodes and
wires so that they are not under the Suction
Cup. If there are already electrodes on the
patient, make sure that they are not under
the Suction Cup. If they are, you must apply
new electrodes.
3. After defibrillation, make sure that the
position of the Suction Cup is correct.
If necessary, adjust the position.
WARNING - CHANGED POSITION
DURING OPERATION
If the position of the Suction Cup changes
during operation or during defibrillation,
immediately push
ADJUST
and adjust the
position. Always use the LUCAS Stabilization
Strap to help secure the correct position.
WARNING - ECG INTERFERENCE
Chest compressions interfere with ECG
analysis. Push
PAUSE
before you start the
ECG analysis. Make the interruption as short
as possible. Push
ACTIVE (continuous)
or
ACTIVE (30:2)
to start the compressions
again.
100901-00 Rev B, Valid from CO J3166 © 2017 Jolife AB