LUCAS
TM
2 Chest Compression System –
Instructions for Use
100901-00 Rev A, Valid from CO J2716 © 2014 Jolife AB
31
9.8 Audible SIGNALS
9.8.1 Audible ALARM SIGNALS, characteristics
NOTE: The ALARM SYSTEM also generates an independent audible ALARM SIGNAL with above
stated sequence of tones by a mechanical buzzer (2400 +/- 50 Hz).
LATCHING ALARM SIGNAL = ALARM SIGNAL that continues to be generated after its triggering
event no longer exists, until stopped by deliberate OPERATOR action.
NON-LATCHING ALARM SIGNAL = ALARM SIGNAL that automatically stops being generated
when its associated triggering event no longer exists.
t
d
= PULSE duration (electrical ON time)
t
s
= PULSE spacing (electrical OFF time)
t
b
= INTERBURST INTERVAL (electrical OFF time)
f
o
= fundamental frequency (first harmonic) of a PULSE
System delays = Sum of alarm signal generation delay and alarm condition delay mean (time from
the occurrence of a triggering event to the generation of its alarm signal).
Audible
signal
name
Sequence of tones
Durations
+/- 5ms
Tone
frequency
+/- 10 Hz
Sound
level
(dB@1m)
+/- 5dB
Situations
System
delays
+/-0.5s
Result
High
priority
alarm
LATCHING ALARM SIGNAL
t
d
= 200ms
t
s
= 100ms
t
s3-4
= 400ms
t
s5-6
= 500ms
t
s8-9
= 400ms
t
b
= 2.5s
f
0
= 530 Hz
f
1
= 1060 Hz
f
2
= 1590 Hz
f
3
= 2120 Hz
f
4
= 2650 Hz
85
Self-test error
during start up.
1 to 10s
Inoperable
device
Compression
pattern outside
limit, too deep
0.6s
Compressions
stop
Compression
pattern outside
limit, too
shallow or
timing failure
30s
Too high
temperature in
LUCAS
0.6s
Internal
hardware error
0.6s
Too high Battery
temperature
0.6s
Compressions
stop
Too low Battery
charge
0.6s
Medium
priority
alarm
NON-LATCHING ALARM
SIGNAL
t
d
= 200ms
t
s
= 200ms
t
b
= 5s
f
0
= 390 Hz
f
1
= 780 Hz
f
2
= 1170 Hz
f
3
= 1560 Hz
f
4
= 1950 Hz
82
Approximately
10 minutes
remaining
operating time
until empty
Battery
Required
action:
Replace Battery
or connect
external Power
Supply
0.6s
The orange
Battery
indication LED
farthest to the
right
illuminates
intermittently
(..)
(...)
(..)
(..)
(..)
(2.5s)
(...)
(2.5s)
(5s)
(5s)
(5s)
(5s) ...
100901-00 Rev B, Valid from CO J3166 © 2017 Jolife AB