Non-invasive Blood Pressure (NBP) Measurement
30
Introduction to the Instrument
NBP Modes
The measurement offers adult, pediatric, and neonatal modes.
The following table lists the cuff inflation limits for each mode:
The following table lists the measurement ranges for each mode:
Safety
The following table lists the maximum limits that ensure patient safety:
If any one of these safety limits is violated, an INOP is generated and the valve opens.
Mode
First Inflation
Subsequent Inflations,
Above Systolic Pressure
Stat Mode
Adult
165
25
15
Pediatric
130
20
15
Neonatal
100
15
15
Mode
Systolic
Diastolic
Mean
Adult
30 - 270
10 - 245
20 - 255
Pediatric
30 - 180
10 - 150
20 - 160
Neonatal
30 - 130
10 - 100
20 - 120
Mode
Maximum
Measurement Time
Maximum Time/
Pressure
Over-pressure
Maximum
Adult
180 seconds
180 seconds for
pressure > 15
mmHg
300 mmHg for > 2
seconds
Pediatric
180 seconds
180 seconds for
pressure > 15
mmHg
300 mmHg for > 2
seconds
Neonatal
90 seconds
90 seconds for pres-
sure > 5 mmHg
150 mmHg for > 2
seconds