Part 2 Isolating and Solving Instrument Problems
Troubleshooting the Instrument
179
MEASSERV
UNSUPPORTD
An incompatible Meas-
urement Server (includ-
ing any M3000A) is
connected to a Monitor
with Rev. E software.
Use only M3001A
Server with a Monitor
with Rev. E software.
SERVERLINK
MALF
Either the current is too
high or the voltage is
too low.
1. Make sure the Server
is connected to the
Monitor.
2. Replace the Server
with a known-good
Server.
3. Check link bar and
cables.
4. Replace the System
Board.
SPEAKER MAL-
FUNCTION
Defective speaker or
cables.
1. Check cable connec-
tions.
2. Replace speaker.
3. Replace the System
Board.
UNSUPPORTED
LAN
An unsupported device
has been connected to
LAN port of the Moni-
tor.
Disconnect the unsup-
ported device.
CO
2
EQUIP MALF
(M3015A and
M3016A)
Malfunction in the CO
2
hardware or the
M1460A transducer
hardware.
Verify failure and
replace the Extension
with a known-good
Extension.
CO
2
NO TRANS-
DUCER
(M3016A)
M1460A Transducer
not connected. Silenc-
ing this INOP switches
the parameter off. If the
transducer is replaced,
the new transducer must
be calibrated.
1. Connect a transducer.
2. Replace a defective
transducer and recali-
brate the parameter.