Installation of Wireless Infrastructure
Installing the Instrument
79
The following cable is required to interconnect the configuring PC to the Wireless LAN
adapter in the M3/M4 patient monitor:
• 9-pin D female - 1/8 in. male stereo phono cable (PN M1360-61675)
Copying the Configuration Tools to the Configuring PC
The first step in the procedure is to copy the configuration tool software from the
IntelliVue
Information Center Application Software CD ROM
to the configuring PC.
Step 1
Turn on the configuring PC to display the
Windows Main Menu
.
Step 2
Insert the
IntelliVue Information Center Application Software CD ROM
into the CD ROM drive of the configuring PC.
Step 3
Open
Windows Explorer
as follows:
a. Click on
Start
in the lower left of the
Windows Main Menu
to open the
Windows Workstation
menu.
b. Click
Programs
to display the
Programs
menu.
c. Click
Windows Explorer
to display folders in the PCs
CD ROM
drive.
Find the
Viridia
directory on the
CD ROM
shown in the figure below and click on
it to display the
tools
directory.
Step 4
Copy the
Viridia\tools\ConfigTool
directory to the drive on the configuring PC
to be used for storing this program as follows:
a. Open the directory to be used on the configuring PC
b. Click on, hold, and drag the
Viridia\tools\ConfigTool
directory to the open PC
directory
NOTE
The
Viridia\tools\ConfigTool
directory file is less than 1.4 Mb so it can be stored on a 1.4
Mb floppy disk for later use, or for transfer to a PC without a CD ROM drive.
Viridia
Содержание M3000A