Part 2 Isolating and Solving Instrument Problems
Troubleshooting the Instrument
195
Recorder Problems (refer also to the Operator’s Manual of the XE-50p
Chart Recorder)
Message
Meaning
Action
“Local Recorder
Malfunction”
prompt message
Command error:
•
Syntax error in either the
Recorder or Monitor
software.
Software problem, not
correctable in the field.
Replace the Recorder.
•
Intermittent connection
to Recorder or faulty
cable.
•
Check cable connection
is secure between the
Monitor and the
Recorder.
•
Check cable, and replace
if necessary.
Recorder hardware fault
detected.
Replace the Recorder.
“Local Recorder
check cable”
prompt message
Communication error:
•
Intermittent connection
to Recorder or faulty
cable.
•
Check cable connection
is secure between the
Monitor and the
Recorder.
•
If connection is sound,
but the problem persists,
replace the cable.
•
Frame error in RS232
protocol (due to a
connection problem or
effects of interference
from external electrical
equipment, such as
flourescent lighting).
•
Check cable connection
is secure between the
Monitor and the
Recorder.
•
If connection is sound,
but the problem persists,
replace the cable.
•
If there is electrical
interference from
external equipment, try to
isolate the cause, and if
possible repair or remove
the cause of the
interference.