Battery Handling, Maintenance and Good Practices
132
Maintaining the Instrument
3
Turn on the monitor, and leave it on until it switches off automatically.
4
Turn the monitor off using the
On/Off/Standby
switch.
5
Reconnect the monitor to the AC power supply.
6
Charge the battery until it is full (the battery LED on the monitor is green).
7
Repeat steps 2 to 6.
A Note on Trickle Charging
Trickle charging means a low current is applied to a fully charged battery to keep the charge
at 100%. (The self-discharge of NiMH batteries is in the range of 20-25% per month!).
Trickle charging accelerates the aging of the battery and is not recommended for batteries
that are only used occasionally. In such cases the battery should be kept separate from the
monitor and charged every three to four months to prevent deep discharge. Deep discharge
occurs when the battery has a relative state of charge of 0%.
NOTE
If you cannot avoid trickle charging, use the Battery Charger and Conditioner (M8043A) to
recondition the battery once a month. See the instructions that accompany the charger for
details on charger use.
Trickle charging only takes place for batteries stored in a monitor connected to AC mains.
You can leave batteries in the Battery Charger and Conditioner (M8043A), as no trickle
charging takes place.
Accessing the Battery Status Window
The Battery Status window shows additional information about the battery. To access this
window, press the Setup key on the front panel and select Battery from the list. The Monitor
displays a screen similar to the following:
The meaning of the messages are described in the following paragraphs.
Time to Empty—This refers to the approximate (+/- 10%) battery-usage time remaining in
the battery. The number is based on the average monitor-loading during the previous minute.
Time to Full—This refers to the approximate (+/- 10%) time remaining to fully charge the
battery. The number is based on the average monitor-loading during the previous minute.
Exit
Capacity Current : 284.0 mAh
Battery Status
designed : 2400 mAh Cycles : 234
remaining : 1247 mAh Voltage : 12.9 V
TimeToEmpty: 112 min +/- 10%(est. 1 min avg)
fullCharge: 1852 mAh Temperature: 23.0 C
Duracell TR36 12.0V SN 00292 07JUL98 SBM 1.0