3 Software Uploads
Checking the Unit for Functionality
30
CO
2
should be approximately 0.03 Vol. %, O
2
should be approximately 20.8 Vol. %, N
2
O and Agent
1 and 2 should be 0 Vol. %. Data valid will have a green LED; ISO accuracy will have a yellow LED
for a successful calibration.
5) In the Status Tab you will see green LEDs for the whole unit except for Breath detection and Breath
Phase: Inspiration.