2 Installation and Patient Safety
Mounting Instructions
22
2
Attach the top mount to the gas analyzer using only the three short countersunk screws supplied
with the mount.
N O T E
Devices installed on the top mount may not weigh more than 12 kg. Make sure that any device
installed on the top mount snaps in properly and is fixed securely to the mount.
Mounting Instructions
N O T E
There are different mounting options available for the IntelliVue G1 / G5. This section covers the
general concepts of safe mount installations and specific steps for the mounting options sold by Philips.
Instructions which ship with a mounting solution should always take precedence over the instructions
described in this chapter. You MUST follow the instructions that ship with the mounting solution,
regardless of manufacturer.
Please mount the IntelliVie G1 / G5 using the Philips Quick Mount solution or another approved
mounting solution. The mounting shall be done in a manner that no patient, operator or other person
can be harmed by a IntelliVue G1 / G5 removed intentionally or released accidentally from the
mount. When using the Quick Mount, be aware of the danger of accidental activation of the Quick
Mount release button when lifting or moving items located under the monitor, such as pole mounts,
etc.
For instructions on how to mount the monitor using the Quick Mount table mount refer to the
Assembly Instructions delivered with the mounting kit M8000-64100 or 453564239731.
WARNING
• It is the customer's responsibility to have the attachment of the mounting hardware to the ceiling,
wall, or mounting rail and the construction of the ceiling, wall, or mounting rail evaluated for
structural integrity and compliance with all local, state and any other required codes by a registered,
professional, structural and/or mechanical engineer.
• Ensure that this commitment has been met before assembling mounts.
• Incorrect mounting and use of inappropriate mounting material may lead to injury. It is the
customer’s responsibility to ensure that the mounting procedures have been performed correctly and
the appropriate mounting devices have been used.