3 Software Uploads
Uploading the Software
38
7) Follow the same steps as with the first file. When the programming process is finished click on the
OK
button.
N O T E
After download and programming is complete, the connection will be lost. You will need to reconnect
in the General Tab in order to check whether the software upload was successful.
8) Please wait a few seconds (approximately 10 sec, maybe select a few Tabs to re-establish the screen)
after reconnection to set the unit to
Operate
and
Disabled for Breath detection
in the General Tab. Then
select the Pneum. Tab again and set the Pump Flow Level to
High
, then select the Zero Tab and select
Auto
again.