4 Testing and Maintenance
Safety Tests
46
The following symbols are used in the diagrams illustrating the safety tests:
The IntelliVue G1 and the IntelliVue G5 are protected against the effects of defibrillation and
electrosurgery.
CAUTION
After each service, maintenance or repair event:
Ensure all fuses accessible from the outside comply with the manufacturer’s specification.
Check:
• the integrity of mechanical parts, internally and externally.
• any damage or contamination, internally and externally.
• that no loose parts or foreign bodies remain in the device after servicing or repair.
• the integrity of all relevant accessories.
Supply mains
Protective earth
L, N
Supply mains terminals
PE
Protective earth terminal
Mains part
Applied part
F-type applied part
Measuring device
Resistance measuring device
Connection to accessible
conductive parts
.........
Optional connection