WARNING
Philips Respironics recommends
using an oxygen monitor that
complies with ISO-21647;
Medical Electrical Equipment
- Particular Requirements for
Basic Safety and Essential
Performance of Respiratory Gas
Monitors.
Flush O
2
Confirmation Screen
WARNING
You should not rely on any
single alarm to detect a circuit
disconnect condition. The Low
Tidal Volume, Low Minute
Ventilation, Low Respiratory
Rate, and Apnea alarms should
be used in conjunction with the
Circuit Disconnect alarm.
FiO
2
You can set the Fractional Inspired Oxygen from 21% to 100% in
increments of 1.
A Flush feature is also available for the FiO
2
setting. A Flush button
appears on-screen if the FiO
2
setting is greater than 21%. Pressing
this button displays the following confirmation screen:
Selecting Yes allows you to temporarily increase the delivered
oxygen concentration to 100% for two minutes. After the two
minutes is completed, the oxygen concentration will return to the
previous FiO
2
setting. Selecting No cancels the action and does not
change the FiO
2
setting.
Circuit Disconnect
This setting enables or disables the circuit disconnect alarm. If
enabled, an audible alarm will sound when a large, continuous air
leak (such as mask removal) has been detected in the circuit.
You can choose Off to disable the alarm. Or, you can increase or
decrease the setting from 5 to 60 seconds in 5 second increments.
For example, a setting of 10 means that the alarm will sound after the
circuit has been disconnected for 10 seconds.
Содержание Trilogy 202
Страница 1: ...User Manual REF 1067236 1067235 JH 12 3 09 ...
Страница 3: ......
Страница 23: ...Trilogy 202 user manual 14 ...
Страница 59: ...Trilogy 202 user manual 50 ...
Страница 75: ...Trilogy 202 user manual 66 ...
Страница 145: ...Trilogy 202 clinical manual 136 ...
Страница 169: ...Trilogy 202 clinical manual 160 ...
Страница 177: ...Trilogy 202 clinical manual 168 ...