Chapter 6
Ventilator Alarms
121
Alarm
Priority
Audible
Visual (Alarm
Indicator Button
and Display)
Device
Action
Clinician Action
Replace
Detachable
Battery
Low or
High,
depending
on cause of
alarm
• •
for Low
• • • • •
for High
“Replace Detachable
Battery” message
appears. If battery
is nearing end of
useful life, message
appears with yellow
background and
button is solid
yellow. If battery fails,
message appears with
red background and
button flashes red.
Continues to
operate
Switch to an alternate
battery or AC power source
while replacing the current
detachable battery.
Ventilator
Service
Recommended
Medium
• • •
Yellow flashing
button;
“Ventilator Service
Recommended”
message
Continues to
operate
If alarm continues, have
device serviced.
AC Power
Disconnected
Medium
• • •
Yellow flashing
button; “AC Power
Disconnected”
message, and a box
appears around
battery in use.
Switches to
alternate
power source
Check batteries and recharge
f necessary. Restore AC power
if available.
Keypad Stuck
Low
• •
Solid yellow button;
“Keypad Stuck”
message.
Continues to
operate
Check keys to determine if
they are lodged in the case.
If alarm continues, place
patient on alternate source of
ventilation and have device
serviced.
Battery
Discharging
Stopped Due to
Temperature
Info
•
“Batt Discharge
Stopped – Temp.”
message
Continues to
operate
If alarm is detected on
the internal battery and
continues after the measures
above have been taken, place
patient on alternate source of
ventilation and have device
serviced.
Battery Not
Charging Due
to Temperature
Info
•
“Batt Not Charging –
Temp.” message
Continues to
operate
If alarm continues after
the measures above have
been taken, place patient
on alternate source of
ventilation and have device
serviced
Battery Not
Charging
Info
•
“Detach Battery Not
Charging” or “Internal
Battery Not Charging”
message
Continues to
operate
If condition continues
for internal battery, place
patient on alternate source of
ventilation and have device
serviced.
Содержание Trilogy 202
Страница 1: ...User Manual REF 1067236 1067235 JH 12 3 09 ...
Страница 3: ......
Страница 23: ...Trilogy 202 user manual 14 ...
Страница 59: ...Trilogy 202 user manual 50 ...
Страница 75: ...Trilogy 202 user manual 66 ...
Страница 145: ...Trilogy 202 clinical manual 136 ...
Страница 169: ...Trilogy 202 clinical manual 160 ...
Страница 177: ...Trilogy 202 clinical manual 168 ...