Chapter 10
System Checkout Procedures
139
Setting
Value
Therapy Mode
S/T
AVAPS (passive circuit only)
Off
IPAP
20 cmH
2
O
EPAP
4 cmH
2
O
FiO
2
45%
Breath Rate
12 BPM
Inspiratory Time
1.6 seconds
Trigger Type (passive circuit
type)
Auto-Trak
Flow Trigger Sensitivity
(Active PAP or Active Flow
circuit type)
6.0 l/min
Leak Compensation (Active
Flow circuit type)
On
Flow Cycle Sensitivity
(Active PAP or Active Flow
circuit type)
20%
Rise Time
1
Ramp Length
Off
All other alarms
Off
B. Options Menu
Modify the settings in the Options menu to match those shown
below in Table 2.
Setting
Value
Menu Access
Full
Detailed View
On
All other settings
Discretionary
C. Turn Device Power On
Press the Start/Stop button on the front of the ventilator. The
system will begin operating using the defined ventilation
settings.
Table 2 - Ventilator Settings
in the Options Menu