126
WARNING
To prevent patient or ventilator
contamination, we recommend
you use a Respironics-approved
main flow bacteria filter (Part
Number 342077) on the patient
gas outlet port. Filters not
approved by Respironics may
degrade system performance.
Installing a New Bacteria
Filter
Preventive Maintenance
Label
Installing the Bacteria Filter
The following illustration indicates where the bacteria filter is
installed on the device.
Preventive Maintenance
The ventilator should be checked and maintained by authorized
service personnel every 12 months at a minimum. Contact Philips
Respironics or an authorized service center for more information.
Refer to the Preventive Maintenance label located on the bottom of
the device for the service date or hours. The label is shown below.
Service Due Date
Number of Blower Hours
When Service is Due
The device’s blower hours are listed in the device software, in the
Information menu.
Refer to the
Trilogy Service Manual
for recommended periodic
maintenance. Recommended periodic maintenance may be based
on blower hours.