Trilogy 202
clinical manual
146
C. Modify Ventilator Alarm Settings
Modify the Low Inspiratory Pressure setting to match the one
shown below in Table 14.
Setting
Value
Low Inspiratory Pressure
6 cmH
2
O
D. Verify Reset
Wait 40 seconds and verify the following auto-reset conditions:
The High Priority audible indicator has stopped
•
sounding
The red light on the
•
Alarm Indicator/Audio Pause
button has stopped flashing
7. Verify the O
2
Blending Operation
This procedure verifies that the Low Oxygen Flow alarm and the Low
Oxygen Inlet Pressure alarm are working properly. These alarms
apply to all circuit types. It assumes that you have attached the
test lung, verify the ventilator settings, and turned on the ventilator
power as described in the Initial Setup section.
Setup
A. Connect an external O
2
Monitor (in accordance with the
recommended manufacture’s guidelines) in-line with the patient
tube. Make sure the O
2
monitor is properly calibrated before
proceeding.
B. Connect the ventilator’s oxygen input port to a source of high
pressure O
2
(50 psi nominal). Turn on O
2
flow to the ventilator.
Ventilator Settings
Set the ventilator FiO
2
setting to 45%.
Table 14 - Modify Low
Inspiratory Pressure Alarm
Setting
Содержание Trilogy 202
Страница 1: ...User Manual REF 1067236 1067235 JH 12 3 09 ...
Страница 3: ......
Страница 23: ...Trilogy 202 user manual 14 ...
Страница 59: ...Trilogy 202 user manual 50 ...
Страница 75: ...Trilogy 202 user manual 66 ...
Страница 145: ...Trilogy 202 clinical manual 136 ...
Страница 169: ...Trilogy 202 clinical manual 160 ...
Страница 177: ...Trilogy 202 clinical manual 168 ...