SELF DIAGNOSTIC FUNCTIONS
This unit has the self diagnostic functions shown below.
Self diagnostic
Situation
Indication
Alarm buzzer
Remote alarm
Sensor abnormality
If the thermal sensor is
disconnected.
Alarm lamp blinks.
E01 and chamber temp. is
displayed alternately.
Intermittent tone.
Alarm status
If the thermal sensor is
short-circuited.
Alarm lamp blinks.
E02 and chamber temp. is
displayed alternately.
If the cascade sensor is
disconnected.
Alarm lamp blinks.
E03 and chamber temp. is
displayed alternately.
If the cascade sensor is
short-circuited.
Alarm lamp blinks.
E04 and chamber temp. is
displayed alternately.
If the filter sensor is
disconnected.
Alarm lamp blinks.
E05 and chamber temp. is
displayed alternately.
If the filter sensor is
short-circuited.
Alarm lamp blinks.
E06 and chamber temp. is
displayed alternately.
If the ambient temperature
sensor is disconnected.
Alarm lamp blinks.
E07 and chamber temp. is
displayed alternately.
If the ambient temperature
sensor is short-circuited.
Alarm lamp blinks.
E08 and chamber temp. is
displayed alternately.
Battery switch
check
When the battery switch is
OFF at the time of alarm
test.
Alarm lamp blinks.
E09 blinks.
---
---
Condenser temp.
abnormality
In the event of failure of fan
motor for cooling the
compressor
E10 and chamber temp. is
displayed alternately.
Intermittent tone.
Alarm status
When an error code (E01 to E10, except for E09) is displayed, contact our sales representative or
agent.
In the case of E09 blinking, turn on the battery switch.