BASIC SCREEN OF CONTROL PANEL
When the power switch is turned on, the basic screen of the control panel is indicated.
1. Display of alarm (Alarm):
Normal condition: “Normal” is displayed.
Alarm-activated, buzzer-delayed: “Alarm” is displayed alternately in reversal/non-reversal character.
Alarm-activated, buzzer-sounding: “Warning” is displayed alternately in reversal/non-reversal character
The complementary message is indicated in the message indication.
●
“Test” is displayed alternately by the reversal/non-reversal character during the alarm test.
2. Display of status (Status):
Normal condition: “Stand-by” is displayed.
The status number is displayed and a complementary message is displayed in the message column when the
operation monitor system detects the specified status. Refer to page 23 for the details.
3. Display of door status (Door):
“Open” is highlighted when the door is open. When the door is
close, "Closed" is displayed . (Normal character)
4. Display of setting (Temp):
Set value of chamber temperature is displayed.
5. Display of present chamber temperature:
Present value of chamber temperature is displayed.
6. Message indication:
Various messages are indicated. Refer to page 23 and 24 for the details.
7. Display of date and time:
The current date and time are displayed.
FUNCTION OF CONTROL PANEL
The following functions are available through control panel:
1. Setting of running operation:
To set a running condition at the start-up. (refer to page 16)
2.
Setting of log cycle and sending to PC:
To set a log cycle of running data and to send a running
log to PC. (page 19)
3.
Setting of date and time:
To set the date and time shown on the basic screen. (page 21)
4.
Setting/display of alarm:
High limit (or low limit) temperature setting(page 16). Power failure alarm
(page 23) and filter alarm (page 23) are indicated.
5. Default setting:
To set the default for LCD panel and communication (DAQ) speed etc. (page 20)
6. Alarm test:
The test of alarm buzzer, alarm lamp and remote alarm is effective by pressing the alarm
buzzer stop key (BUZZER) for about five seconds during normal operation. Pressing the key again
finishes the alarm test.
1
2
4
5
7
0
1
0
2
1
1
1
2
1
3
1
4
1
5
1
6
1
7
1
8
1
9
2
0
2
1
2
2
2
3
2
4
2
5
2
6
2
7
2
8
2
9
3
0
3
1
3
2
3
3
3
4
3
5
3
6
3
7
3
8
3
9
4
0
1
T o p S c r e e n
2
T e mp - 1 5 0
o
C
3
A l a r m
:
N o r m a l
4
S t a t u s
:
S
t a
n
d
- b
y
5
D o o r
:
C
l o
s e
d
6
2 0 0 6 / 0 7 / 0 1 1 2 : 0 0 : 0 0
-150
℃
3
6
Содержание MDF-C2156VAN Series
Страница 2: ......