   PRECAUTIONS FOR SAFE OPERATION 

 
 
 
 
 

Do not use the unit outdoors.

    Current leakage or electric shock may result if the unit is exposed to 

rain water.

 

 

Only qualified engineers or service personnel should install the unit.

  The installation by 

unqualified personnel may cause electric shock or fire. 
 

Install the unit on a sturdy floor and take an adequate precaution to prevent the unit from 
turning over.

    If the floor is not strong enough or the installation site is not adequate, this may result 

in injury from the unit falling or tipping over.

 

 

 
Never install the unit in a humid place or a place where it is likely to be splashed by water.

   

Deterioration of the insulation may result which could cause current leakage or electric shock.

 

 
Never install the unit in a flammable or volatile location.

 This may cause explosion or fire.

 

 

Never install the unit where acid or corrosive gases are present

 as current leakage or electric 

shock may result due to corrosion.

 

 

 

Always ground (earth) the unit to prevent electric shock.  

If the power supply outlet is not 

grounded, it will be necessary to install a ground by qualified engineers. 

 

Never ground the unit through a gas pipe, water main, telephone line or lightning rod.

    Such 

grounding may cause electric shock in the case of an incomplete circuit. 

 

Connect the unit to a power source as indicated on the rating label attached to the unit.   

Use 

of any other voltage or frequency other than that on the rating label may cause fire or electric shock. 

 

Never store volatile or flammable substances

 in this unit if the cylinder cannot be sealed.    These 

may cause explosion or fire.

 

 

Do not insert metal objects such as a pin or a wire into any vent, gap or any outlet

 

on the unit.

   

This may cause electric shock or injury by accidental contact with moving parts.

 

 

Use this unit in safe area when treating the poison, harmful or radiate articles.

  Improper use 

may cause bad effect on your health or environment. 
 

Turn off the power switch (if provided) and disconnect the power supply to the unit prior to any 
repair or maintenance 

of the unit in order to prevent electric shock or injury.

 

 

Do not touch any electrical parts (such as power supply plug) or operate switches with a wet 
hand.

    This may cause electric shock. 

 
 
 

WARNING

 

Содержание MDF-C2156VAN Series

Страница 1: ...ies Please read the operating instructions carefully before using this product and keep the operating instructions for future use See page 40 for all model numbers Operating Instructions Ultra Low Tem...

Страница 2: ......

Страница 3: ...NU Tools P 20 Initialization Tools Default Setting P 20 Setting of date time log interval Tools Date Time P 21 Setting of key lock password Tools Key lock PW Setting P 22 ALALRM BUZZER P 22 MONITOR OF...

Страница 4: ...ating instructions are subject to change without notice for improvement of performance or function Contact our sales representative or agent if any page of the operating instructions is lost or the pa...

Страница 5: ...an instruction must be followed Be sure to keep the operating instructions in a place accessible to users of this unit Label on the unit This mark is labeled on the cover in which the electrical compo...

Страница 6: ...wer supply outlet is not grounded it will be necessary to install a ground by qualified engineers Never ground the unit through a gas pipe water main telephone line or lightning rod Such grounding may...

Страница 7: ...the power supply outlet grip the power supply plug not the cord Pulling the cord may result in electric shock or fire by short circuit Disconnect the power supply plug before moving the unit Take care...

Страница 8: ...nces such as acid or alkali in this unit if the cylinder cannot be sealed These may cause corrosion of inner components or electric parts Check the setting when starting up of operation after power fa...

Страница 9: ...40 o C Maximum relative humidity 80 for temperature up to 31 o C decreasing linearly to 50 relative humidity at 40 o C Mains supply voltage fluctuations up to 10 of the nominal voltage Transient overv...

Страница 10: ...8 FREEZER COMPONENTS Rear side Inside of side table 16 27 10 5 9 17 15 18 19 20 22 23 21 26 25 24 6 1 3 2 5 7 9 8 10 11 12 13 14 16 17 4 27...

Страница 11: ...rm cable output port in the inside 11 Panel door lock To lock the control cover to avoid the setting by accidental contact 12 Panel door There is a controller such as a power switch inside the panel d...

Страница 12: ...WER Power switch of the freezer 25 Battery switch It is the switch of the battery for the power failure alarm Turn it on usually Turn it off when you do not drive the freezer for a long time More than...

Страница 13: ...ckup cooling kit is activated 5 Figure input key User for operation setting 6 Menu button MENU To open the menu window 7 Shift key Upward downward rightward leftward To move the cursor on the LCD pane...

Страница 14: ...n or noise WARNING Install the unit on a sturdy floor If the floor is not strong enough or the installation site is not adequate this may result in injury from the unit falling or tipping over Select...

Страница 15: ...the leveling foot Stretch the leveling feet by rotating them to contact them to the floor Ensure the unit is installed horizontally 3 Fixing the unit Two fixtures are attached to the rear of the fram...

Страница 16: ...ushed again an alarm test movement is finished 8 Push the backup test switch and confirm that liquid N2 flows into the freezer chamber 9 Put the sample to cryopreserve into the freezer chamber Turn of...

Страница 17: ...yed 6 Message indication Various messages are indicated Refer to page 23 and 24 for the details 7 Display of date and time The current date and time are displayed FUNCTION OF CONTROL PANEL The followi...

Страница 18: ...rature 5 o C 20 o C The factory setting is 10 o C Low Alarm It is the set point of the low temperature alarm Temperature settable range Chamber temperature 5 o C 20 o C The factory setting is 10 o C R...

Страница 19: ...ure setting screen Temp Setting input 0 in the Key Lock field and press the enter key ENTER 2 The cursor is moved to the password column Password Input 4 digit password and press the enter key ENTER T...

Страница 20: ...mperature 50 o C 25 o C 25 o C 100 o C 100 o C 175 o C It is moved with and to date Past date New date A data interval is 2 minutes from 30 minutes The factory setting is 15 min With 15 minutes interv...

Страница 21: ...ion starts when a transfer a capture of the text and a preservation file name are specified and MENU Start is selected 2 The transmission is finished with Finished indication PC setting As for setting...

Страница 22: ...p of the back light 1 Blue 2 White DAQ Speed should use 0 2400 It is DAQ standard command mode DAQ ID select any ID between 1 and 250 when an optional interface beard is attached DAQ Mode 0 setting ch...

Страница 23: ...is set up with MENU OK ENTER Door Delay The settable range of the door alarm delay time is 1 minute and 15 minutes The factory setting is 2 minutes Log Interval Settable between 2 minutes and 30 minu...

Страница 24: ...arm buzzer sounds again if an alarm is not solved in the time when it is set up with Ring Back When alarm sounds a remote alarm is activated too Door alarm buzzer intermittent tone It is informed with...

Страница 25: ...TUS indicated is to cope with it ALARMS SAFETY FUNCTIONS This product has the alarm safety function of the table 2 Table 2 Alarm safety function list Alarm safety Situation Indication Buzzer Safety op...

Страница 26: ...disconnected Message indication Error 05 Filter Sensor is Open Intermittent tone Remote alarm If the filter sensor is short circuited Message indication Error 06 Filter Sensor is Short Intermittent to...

Страница 27: ...L The analog output terminals can output the chamber temperature as a voltage signal The output range is between 0 mV and 100 mV Range Output Temp 160 o C 40 o C 0 5 mV o C Example Output at 150 o C 1...

Страница 28: ...should be done by high pressure gas works specialists The pressure of liquid N2 cylinder should be regulated at 49 0 kPa G 0 5 kgf cm2 G 2 Operate the freezer until the freezer temperature reaches the...

Страница 29: ...electric insulation and cause failure The compressor and other mechanical parts are completely sealed This unit requires absolutely no lubrication Check the backup cooling kit by pressing test switch...

Страница 30: ...ther freezer or container which contains liquid N2 or dry ice Switch off the remote alarm and backup cooling kit Switch off the power supply 2 Open the door and remove the inner lid Leave the freezer...

Страница 31: ...engineer or service personnel only to prevent the electric shock Removal of lead storage battery 1 Turn off the power switch battery switch and disconnect the power supply plug 2 Lift the handle of th...

Страница 32: ...it is left in these cases The cooling is poor The environmental temperature is too high The door is not shut tightly The inner lid is not installed correctly The set temperature is not set properly T...

Страница 33: ...s procur est aliment par une batterie au plomb tanche Apr s la fin de la vie utile de la batterie en vertu de diverses r glementations gouvernementales et locales il est ill gal de l liminer avec les...

Страница 34: ...ion und L nder mit Recyclingsystemen Dieses Symbol auf den Produkten der Verpackung und oder den Begleitdokumenten bedeutet dass gebrauchte elektrische und elektronische Produkte sowie Batterien nicht...

Страница 35: ...ive au produit chimique concern Spanish Eliminaci n de Aparatos Viejos y de Pilas y Bater as Solamente para la Uni n Europea y pa ses con sistemas de reciclado Estos s mbolos en los productos su embal...

Страница 36: ...Neste caso estar em conformidade com o estabelecido na Diretiva referente aos produtos qu micos em causa Italian Smaltimento di vecchie apparecchiature e batterie usate Solo per Unione Europea e Nazi...

Страница 37: ...l Onderstaand symbool Dit symbool kan in combinatie met een chemisch symbool gebruikt worden In dit geval volstaan de eisen die zijn vastgesteld in de richtlijnen van de desbetreffende chemische stof...

Страница 38: ...ion procedure The installation method of MTR 155H 1 Remove four screws shown by arrows on the front After that remove a temperature recorder cover 4 screws 2 Remove the recorder sensor cover 2 screws...

Страница 39: ...ing MDF S30150 and temperature recorder fitting to the mounting space and fix it to the front side of the front panel Fix with four screws 8 Check that setup temperature is the same as chamber tempera...

Страница 40: ...of the TERMINATOR switch of the interface board MTR 480 Fig 3 4 Connect to the connector side the back side of the interface board of a communication cable RS 232C or RS 485 to use for the interface b...

Страница 41: ...larm Filter check Remote alarm contact Allowable contact capacity DC 30 V 2 A Battery Lead storage battery DC 6 V 7 2 Ah x 4pcs Auto recharge Accessories 1 set of key 1 scraper 1 set connect tube for...

Страница 42: ...ambient temperature 30 o C no load Temperature control range 125 o C to 150o C ambient temperature 30 o C no load Power source 220 V 220 V AC 230 V 240 V Power source 50 Hz 60 Hz 50 Hz Rated power con...

Страница 43: ...unit Unit interior No contamination Decontaminated Contaminated Others 3 Instructions for safe repair maintenance disposal of the unit a The unit is safe to work on b There is some danger see below Pr...

Страница 44: ...Printed in Japan LDCL028100 1 S0418 10418 PHC Corporation 2018 1 1 1 Sakada Oizumi machi Ora gun Gunma 370 0596 Japan...