21
VARIOUS SETTING (MENU/Tools)
Setting of date, time, log interval(Tools/Date Time)
1.
On the setting screen (Select Tools), select “Data Time”, and press the enter key (ENTER). A date,
time and a log interval setup screen (Data Time) is indicated. Set up each parameter.
Each parameter setting range:
●
July 1st, 2006 12:00:00 is set.
It is input with 060701 in the Date cell.
It is input with 120000 in the Time cell.
It is set up with MENU/ OK (ENTER).
●
Door Delay: The settable range of the door alarm delay time is 1 minute and 15 minutes. (The factory
setting is 2 minutes)
●
Log Interval: Settable between 2 minutes and 30 minutes. (Factory setting is 15 minutes.) With 15
minutes interval, recording for about 5 weeks is available.
●
Comp Delay: Delay time for high temperature side/low temperature side compressor after power failure.
The settable range is between 2 minutes and 15 minutes. (Factory setting is 2 minutes.)
2.
After setting, press the menu button (MENU), select “OK” and press the enter key (ENTER). The
setting is memorized.
2