67
14.
After dry heat sterilisation is completed, buzzer
sounds and the
Sterilisation Step6
screen is
displayed. If dry heat sterilisation finishes
successfully, “Sterilisation Finished Successfully.” is
displayed. Press the
OK
key to return to the
Top
screen.
Note:
If dry heat sterilisation ends in failure by
something abnormal during dry heat sterilisation,
“Sterilisation Stopped With Error.” is displayed.
Resolve the cause of the error displayed in the
message display field, in reference to Table 8 on
page 90~91.
Press the
OK
key to return to the
Top
screen, and
perform dry heat sterilisation again.
Note:
This screen may be displayed when the dry
heat sterilisation temperature is between 160
o
C~180
o
C due to low ambient temp. and low
voltage although sterilisation has been successful.
Press the
OK
key to return to the
Top
screen.
15.
Open the outer and inner doors and place all the attachments back into the chamber. Pour sterile
distilled water in the humidifying pan. Start cultivation after both of the chamber temperature and the CO
2
density is stable at the set value.