DRY HEAT STERILISATION
●
Step3 is the process to heat the inside of the
chamber to 180
o
C (heating process).
·
After the entire inside of the chamber exceeds
180
o
C, the display leads the
Sterilisation Step4
screen. The displayed temperature comes to more
than 180
o
C sometimes.
●
Step4 is the process to sterilise the inside of the
chamber by dry heat by keeping the chamber
temperature to 180
o
C or higher for 60 minutes
(sterilisation process).
Note:
By pressing the
Abort
key in step3 or step4,
dry heat sterilisation is stopped in the middle of dry
heat sterilisation and goes to Step5 (the cooling
process). After that, the display leads the
Sterilisation Step6
screen (“Sterilisation Stopped
Manually.” is displayed). Press the
OK
key to return
to the
Top
screen.
●
Step5 is the process to cool the inside of the
chamber to 40
o
C (cooling process).
Note:
·
It is not possible to change the setting of 40
o
C.
·
It is not possible to operate the
Abort
key during
the cooling process.