INCUBATOR COMPONENTS
LCD touch panel
The following display (called the
Top
screen) will appear when the power switch is turned ON.
Note:
It takes approximately 20 seconds until the
Top
screen is displayed. During warming-up, “Status:
Gas sensor initializing” is displayed in the Message display field (
13
), and “--.-” is displayed in the Present
CO
2
density display field (
4
).
2 6
7 8
3
1
1. Present temperature display field
The present chamber temperature is displayed.
2. Set temperature value display field
The set value of chamber temperature is
displayed. Default setting: 37
o
C.
3. Heating indicator
This indicator lights when the heater is energized.
4. Present CO
2
density display field
The present chamber CO
2
density is displayed.
Nothing is displayed when CO
2
density is set 0 %.
5. Set CO
2
density value display field
The set value of the chamber CO
2
density is
displayed. Default setting: 0 %.
6. CO
2
gas injection indicator
This indicator lights when CO
2
gas is being
injected.
7. CO
2
gas supply line indicator A and B*
The CO
2
gas supply line (connecting port for
CO
2
gas pipe) used now is displayed. When
the CO
2
gas automatic changer function
changes the empty CO
2
gas cylinder to the
other, the empty indicator is displayed in
reverse video and blinks.
8.
CO
2
gas supply line select
key*
This is a key to select CO
2
gas supply line A
or B (Connecting port A or B for CO
2
gas
pipe). When an optional gas auto charger
MCO-21GC is installed, CO
2
gas supply line
A/B changes over automatically when CO
2
gas cylinder is empty. Changeover is also
workable by pressing this key.
* Only when an optional component MCO-21GC (Gas auto charger) is installed, this key is workable. They are
not displayed when the MCO-21GC is not installed.
5