13
10 11 9 12
13
18 17 16 15
9. Present date/time display field
Present date and time are displayed. The
date and time are simply set when the
incubator is shipped from the factory. Refer to
page 60~61 for details.
10. Over heat display
High limit temperature alarm is activated:
“Over Heat” is displayed alternately in normal
characters and reverse video.
11. UV lamp condition display
UV lamp ON: “UV : On” is displayed.
UV lamp OFF: “UV : Off” is displayed.
Note:
Nothing is displayed when an optional
UV System Set For Dry Heat MCO-170UVSD
is not installed to the MCO-170AICD.
12. Outer door (opening/closing) display
Open: “Door : Open” is displayed alternately
in normal characters and reverse video.
Close: “Door : Closed” is displayed.
Locked: “Door : Locked” is displayed.
13. Message display field
Alarms, errors or messages are displayed
when fault occurs. Refer to page 88~91.
Note:
When there are a number of
alarms/errors, the display shows the message.
For example, if 2 alarms/errors occur in total,
the display shows “1/2”.
14.
Message select
key
When there are a number of alarm/errors,
press this key to change the displayed
message in the Message display field.
15.
Menu
key
Press this key to lead the
Menu
screen. It is
possible to set various setting on the
Menu
screen. Refer to page 30.
16.
Sterilisation
key
This key is to perform dry heat sterilisation.
Refer to page 63~67.
Dry heat sterilisation temp. and time: 180
o
C -
60 min.
14