56
OPERATION/ALARM/STERILISATION LOG
4.
On the
Sterilisation Log
screen, the latest 6
sterilisation logs are displayed.
Note:
When 7 or more sterilisation logs exist, by
pressing the top (
▲
) or the bottom (
▼
) log, the log
table scrolls one by one and hidden sterilisation
logs can be seen.
·
Press the
Back
key to return to the
Log
screen.
·
Press the
Top
key to return to the
Top
screen.
5.
On the
Sterilisation Log
screen, by inputting the
number of logs into the Last XX Logs input box, the
latest sterilisation logs of specified number of logs
are displayed.
Settable range: 1 log~250 logs.
Note:
The error of about 1 minute may be observed
during 1 month. Refer to page 60~61 for the
procedure of setting time.
·
Press the
Back
key to return to the
Log
screen.
·
Press the
Top
key to return to the
Top
screen.
●
On the
Sterilisation Log
screen of procedure
4
or
5
, sterilisation log data can be exported in CSV format
to the USB memory inserted into the USB port.
6.
Insert the USB memory into the USB port.
Note:
It is not possible to use a USB memory with security functions that requires entering password.
7.
Press the
Export
key.
8.
When the export is complete, Information dialog
box is displayed. Press the
OK
key. Refer to page
58 for the details about abnormal export or
exported file name.
9.
Press the
Top
key to return to the
Top
screen.