Section 4-10-3
<Mounting>
Follow the removing procedure in reverse.
Note 1: Before attaching Dust cover, confirm that the cable from
the lamp shall be routed as shown in right figure.
Note 2: When connecting a cable to Cover open sensor, a cable
band shall be placed on U-shape slot of the frame. This
cable setting is important to avoid the cable be
disconnected by pulling the cable at assembling the
scanner later. Long-nose pliers may be necessary to
connect the cable to Cover open sensor.
Cable
U-shape slot of frame
Cover open sensor
Cable band